Appendix C. Specifications
150
AquaTrans™ AT600 User’s Manual
Note:
As indicated on the previous page, the current version of your AT600 firmware must be 01.02.23 or later. If your
version is older than this, you cannot use this update method. Software versions 01.02.24, 01.03.xx, 02.xx.xx are
examples of acceptable versions.
C.2.2 Update steps
If your AT600 firmware version qualifies for the field firmware update process, complete the following steps:
1.
Prepare the RS485 Modbus connection:
a.
Disconnect the main power from the AT600.
b.
Wire the Modbus connection as described in “Wiring the Modbus communication” on page 24.
2.
Locate the AquaTrans flow meter Software update (version 20161117V1.2 or later) software on your PC. If the folder
containing the software is zipped, you must unzip it prior to use.
3.
Run the update software by clicking on the upgrade.exe file (see figure 37 below). Installation of the software on
the PC is not required.
Figure 37: Running the update.exe software
4.
Click on the COM port button and enter the specific com port on the PC that is connected to the AT600 Mod
-
bus/service port.
5.
Click on the open image file button, and open the Panametrics image file to be used for the AT600 firmware
update.
6.
Click on the start To upgrade button. After verifying that the selected image file and Com port are correct, click
the OK button to start the update process.
7.
After the progress bar indicates that the firmware update is 100% complete (about 10 minutes), the message
shown in figure 38 below will be displayed. Note that the AT600 will automatically reboot in 30 seconds.
Figure 38: Reboot message
8.
After the reboot is complete, refer to “Check the current firmware version” on page 149 and confirm that the new
firmware version has been installed. If the original firmware version is still installed, the AT600 was not able to val
-
idate the image file that was used - contact Panametrics for assistance.
Содержание Panametrics AquaTrans AT600
Страница 1: ...panametrics com BH001C11 Rev F AquaTrans AT600 User s manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii AquaTrans AT600 User s manual no content intended for this page...
Страница 20: ...Chapter 1 General installation instructions 10 AquaTrans AT600 User s manual no content intended for this page...
Страница 36: ...Chapter 3 Wiring the AT600 electronics 26 AquaTrans AT600 User s manual no content intended for this page...
Страница 84: ...Chapter 4 Initial setup and programming 74 AquaTrans AT600 User s manual no content intended for this page...
Страница 150: ...Chapter 6 Communication 140 AquaTrans AT600 User s manual no content intended for this page...
Страница 158: ...Appendix B Specifications 148 AquaTrans AT600 User s Manual no content intended for this page...
Страница 162: ...Appendix C Specifications 152 AquaTrans AT600 User s Manual no content intended for this page...
Страница 164: ...Appendix D Menu maps 154 AquaTrans AT600 User s Manual D 2 The main menu Figure 42 The main menu...
Страница 172: ...Warranty 162 AquaTrans AT600 User s Manual...
Страница 173: ......