Copyright 2007 Baker Hughes Company.
English–DPI 610/615 IS Instruction Manual | 1
General
1. Introduction
1.1 General
The DPI
610
IS and DPI
615
IS intrinsically safe instruments measure and display pneumatic and
hydraulic pressure applied to the test port.Pressure measurement can be absolute, gauge and
sealed gauge and in ranges from 2.5 mbar to 700 bar (1.0 inH
2
O to 10000 psi).
Calibrator versions of this instrument contain pneumatic or hydraulic pressure generation
components to produce pneumatic pressure ranges between -1 to 20 bar (-14.5 psi to 300 psi)
and hydraulic pressure ranges up to 400 bar (6000 psi).
Using external electrical connections, the DPI
610
IS and DPI
615
IS intrinsically safe
instruments measure ±30 V dc and ±55 mA. An integral sensor provides measurement of ambient
temperature. Additional sensors (option B1) connect to an external connector and extend the
pressure measurement range and include differential pressure measurement. The DPI
615
instrument has an RS232 connector to enable downloading of test data to a compatible
documenting system. Six alkaline C size batteries, IEC Type LR14, power the instrument.
1.2 Description of Procedures
The procedures apply to both the DPI 610 IS and the DPI 615 IS instruments unless otherwise
stated. In the procedures in this manual, hard (fixed function) and soft (variable function) key
operations are shown in bold type:
TASK
and
F1
. These statements mean press the
TASK
key
and press the
F1
key. Soft key operations can be assigned to both the
F1
and
F2
keys. Where a
specific soft function is referred to it is written in bold italics (e.g.)
PROCESS
.
This instrument has a number of operating modes that are described in simplified form in the
following sections. Diagrams accompanying the procedures give typical selection sequences and
shaded controls indicate that this control key should be pressed in the appropriate sequence.
Diagrams should be read from left to right, top to bottom where appropriate. A shaded display soft
box indicates that the function key immediately below that soft box should be pressed (either
F1
for the left hand soft box or
F2
for the right).
In the above diagram the following key sequence is indicated.
a. Press the F2 key (the key immediately below the
UNITS
soft box).
b. Use the
Up
and
Down
cursor keys (only) to select the required option. (If all keys shaded,
use all these keys to select or enter data).
c. Press the
ENTER
key.
INFORMATION
Zinc-carbon and zinc-chloride cells must not be used in this
instrument. Use only the battery types as shown in “Batteries” on page
SELECT UNITS
TASK : BASIC
VOLTAGE
PRESSURE INT
CURRENT
PRESSURE
UNITS
SELECT UNITS
OF PRESSURE
Содержание Druck DPI 610 IS
Страница 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Страница 2: ......
Страница 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Страница 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Страница 20: ...Copyright 2007 Baker Hughes Company 10 DPI 610 615 IS Instruction Manual English Chapter 2 Getting Started...
Страница 38: ...Copyright 2007 Baker Hughes Company 28 DPI 610 615 IS Instruction Manual English Chapter 5 Advanced Task...
Страница 64: ...Copyright 2007 Baker Hughes Company 54 DPI 610 615 IS Instruction Manual English Chapter 8 Calibration...
Страница 75: ......