iSonic 4000

Open-Channel Flow Meter

HYB-UM-02509-EN-04 (November 2018)

User Manual

Содержание iSonic 4000

Страница 1: ...iSonic 4000 Open Channel Flow Meter HYB UM 02509 EN 04 November 2018 User Manual...

Страница 2: ...iSonic 4000 Open Channel Flow Meter Page ii November 2018 HYB UM 02509 EN 04...

Страница 3: ...gs 7 Installation 8 Installation the EchoPod DL 10 Sensor 8 Mounting Positions 9 Power Connections 9 Auxiliary Power 10 Operation 12 Function Buttons 12 Display Icons 12 Initial Screens 12 Setting a P...

Страница 4: ...29 Meter Setup 29 Measurements 29 Inputs Outputs 30 Total 30 Communications 30 Miscellaneous 30 Info 30 Pin 30 Login 30 Flow Meter ModBus Register Table 31 iSonic 4000 Flow Meter Conversion Table 35 R...

Страница 5: ...mains before removing any covers Power Connection Use only the type of power source suitable for electronic equipment If in doubt contact your distributor Ensure that any power cables are of a suffici...

Страница 6: ...n IP67 device in a robust wall mounted metal case with a large graphic display The flow meter menu is operated with three front panel high endurance buttons The flow meter is powered externally by 92...

Страница 7: ...System Settings Flow Meter Tool Settings Settings Control Panel Driver Details System Description Page 7 November 2018 HYB UM 02509 EN 04...

Страница 8: ...gasket onto the threaded end of the sensor 2 Screw the sensor into the stainless steel mounting bracket NOTE N Install the sensor at a maximum of 49 21 in 125 cm above the flume bottom minimal measur...

Страница 9: ...nsor MT V H in DN g sec l sec in mm in mm 3 75 14 26 54 30 71 780 12 00 305 6 150 30 12 114 30 71 780 15 98 406 9 230 77 67 284 38 19 970 22 52 572 12 305 157 98 598 contact factory contact factory 18...

Страница 10: ...T BREAKER 1 Slightly loosen the lower cover screws 2 Completely loosen both upper cover screws 3 Open the cover to the lower side 4 Push the auxiliary power cable through the upper cable inlet 5 Conne...

Страница 11: ...n ON or OFF For the RS485 connect the A wire to the Y terminal and the B wire to the Z terminal 422 232 485 A RxD B Z TxD B Y A G GND USB USB Device CDC Host Mass Storage Micro USB Ethernet Ethernet I...

Страница 12: ...ce error 0 Sensor not connected No keyword active M Sensor measuring USB active S Simulation active Meter Setup Main Menu Menu Header Scroll Bar Indicates a Submenu Submenu Initial Screens From the Ma...

Страница 13: ...ss the right arrow to highlight ON or OFF 5 With either ON or OFF highlighted press the up arrow to display ON 6 Press EXIT SAVE to save the ON setting 7 With the Control menu highlighted press the up...

Страница 14: ...tputs Totalizer Reset Communication Miscellaneous Information Pin The security levels are A Administrative S Service U User Parameters indicated by the battery icon if changed will affect battery perf...

Страница 15: ...unit is powered from battery WarmUpTime S Powering time of sensor s before measurement larger interval is set when unit is powered from battery LowerRangeValue A The minimum level value of used senso...

Страница 16: ...r m3 s Cubic Meters Sec MG d MillionGallons Day m3 min Cubic Meters Min IG s ImperialGallons Sec m3 h Cubic Meters Hour IG min ImperialGallons Min ft3 s Cubic Feet Sec IG h ImperialGallons Hour ft3 mi...

Страница 17: ...anh Flume6 Manhole Flume 6 Manh Flume8 Manhole Flume 8 Manh Flume10 Manhole Flume 10 Manh Flume12 Manhole Flume 12 Equation Params A Exponent value in for equation Q K h exp Exponent Coefficient value...

Страница 18: ...lume 36 in Q 2 1900 h1 57 0 925 Parshall flume 48 in Q 2 9600 h1 58 0 925 Parshall flume 60 in Q 3 7500 h1 59 0 925 Manhole flume 4 in Q 0 2343 h1 95 0 149 Manhole flume 6 in Q 0 3026 h1 95 0 227 Manh...

Страница 19: ...s n Gauckler Manning coefficient s 3 m L Width m h Water level m a Angle I Water surface slope m m Manning pipe a sina r 2 3 I a sina r2 where Equation Q 1 n 2 a 2 a 2 2 arcsin 2 h r h2 h r r 2 arcsin...

Страница 20: ...5 mA and 23 mA OFF Analog signal is based on flow rate and always within the configured range 3 5 mA During alarm conditions the analog signal is 3 5 23 mA During alarm conditions the analog signal i...

Страница 21: ...pulses per totalizer unit via pulse scale Digital Outputs 1 and 2 S The two outputs can be operated as open collector passively or actively Passive output Active output if analog output is not used S...

Страница 22: ...volume or open channel flow rate on page 23 Error alarm Indicates when the meter has error an condition Pump Control Starts or stops the pump See Figure 3 Tank volume or open channel flow rate on page...

Страница 23: ...Figure 3 Tank volume or open channel flow rate Clear Total Total A Resets the totalizer within the ClearTot item on the Flow Meter Tool software Programming Page 23 November 2018 HYB UM 02509 EN 04...

Страница 24: ...Power up The number of times that the unit has been powered on Language The unit supports these languages English German Czech Spanish French Russian Date Set the system date in the format DD MM YY us...

Страница 25: ...but cannot change them Control Activate and deactivate the PIN User User logged in with this PIN can access all User levels but do not have access to Service or Admin functions Service User logged in...

Страница 26: ...LED 7 LED 6 LED 8 LED 13 LED 5 Display Board The following LEDs on the board control the operation of the device LED1 No function attached LED3 Communication transmit On active LED5 Flash memory acti...

Страница 27: ...tivity Pulse length Configurable up to 2000 msec Datalogger 2 MB capacity with 130 000 logged lines date level flow rate tank volume Display Graphical LCD 64 128 backlight actual flow rate totalizers...

Страница 28: ...00 flow computer 2 10 in 53 29 mm 0 38 in 9 7 mm R3 R3 1 00 in 25 3 mm 11 11 1 34 in 34 0 mm 4 00 in 101 50 mm 0 43 in 11 00 mm 0 38 in 9 60 mm 0 43 in 11 00 mm 0 38 in 9 60 mm 0 44 in 11 20 mm 2 24 i...

Страница 29: ...lection Table Exponential Eq Contract Weir Suppress Weir Cipoletti Weir V NotchWeir30 V NotchWeir45 V NotchWeir60 V NotchWeir90 Manning Rect Manning Pipe Pars Flume 1 Pars Flume 2 Pars Flume 3 Pars Fl...

Страница 30: ...0 90 80 70 60 50 40 30 20 10 0 0 Total Total Clear Tot Exit Communications Modbus MODBUS Address RS 232 422 485 Baud Rate 1200 2400 4800 9600 19200 38400 115200 Parity Even Odd Ethernet Received Pack...

Страница 31: ...0 10 mA 0x00A1 U16 Service OUT1_LOW Digital Output setting 0x00A2 U16 Service OUT1_HIGH Digital Output setting 0x00A3 U16 Service OUT1_MODE 0 normally open 1 normally closed 0x00A4 U16 Service OUT1_OP...

Страница 32: ...INTA Dry calibration 0x0128 Float Factory CURRENTLOOP_POINTB Dry calibration 0x012A U16 Service SIMULATION Not stored in non volatile memory 0 0 0 10 10 0 20 20 0 30 30 0 40 40 0 50 50 0 60 60 0 70 70...

Страница 33: ...tting 0x0268 U16 Service MOVING_AVERAGE Filter setting 0x0279 Float Read only ANALOG_OUTPUT_K Dry calibration 0x0281 Float Read only ANALOG_OUTPUT_Q Dry calibration 0x02B3 Float Service ANALOG_OUTPUT_...

Страница 34: ...24 21 Parshall Flume 36 22 Parshall Flume 48 23 Parshall Flume 60 24 Manhole Flume 4 25 Manhole Flume 6 26 Manhole Flume 8 27 Manhole Flume 10 28 Manhole Flume 12 29 Exponential Equation 0x0308 Float...

Страница 35: ..._ COEFFICIENT Open channel calibration 0x0404 Float Admin OPENCHANNEL_WIDTH Open channel calibration 0x0406 Float Admin OPENCHANNEL_ANGLE Open channel calibration 0x040C Float Admin OPENCHANNEL_RADIUS...

Страница 36: ...ght to change product or system specifications without notice except to the extent an outstanding contractual obligation exists 2018 Badger Meter Inc All rights reserved Control Manage Optimize The Am...