7
If any of the transducers reads a
number that is not close to zero, click
OK, and then choose the menu item
Setup / Transducers / Zero
Calibration to zero the transducers.
Make sure that the facepiece has
been removed from the PosiChek
test head prior to zeroing the
transducers.
↓
↓
Note: A transducer stability test will
automatically be performed following
the transducer zero calibration.
Once the zero calibration has been
completed, click OK and return to the
diagnostics screen.
If each of the transducers now reads
a number that is close to zero, click
OK and return to testing. If the
problem persists, continue to step 4.
4. From the Setup / Transducers /
Diagnostics screen, click on the
Actions button followed by “Show
Details” to view the raw counts of
each transducer.