Pur•Chek

 Pro

Instruction 2100-9001

14

5.3  Changing the White Plastic Sample Filter

When inspecting the sample fi lter, look completely around the entire outside diameter of the white fi lter element 
located inside of the clear plastic housing.  Look for red spots or the beginnings of discoloration on the white 
outside diameter of the element.  Do not look into the round ends of the white element for red spots or discolor-
ation.  The round ends of the fi lter may always appear red.  If red spots or discolorations are discovered on the 
outside diameter, the sample fi lter requires replacement to prevent the infl ux of particulate and oil mists into the 
instrument.

Obtain a replacement fi lter, P/N 2100-0006.  Remove the existing fi lter from the retaining clip of the instrument 
by pulling straight up and out.  CAREFULLY remove the fl exible, black rubber tubing connections from both 
ends of the existing fi lter.  DO NOT allow the tubes to slip back into the internal portion of the case. Discard the 
existing fi lter in an environmentally friendly manner.

Install the tube ends onto the barbs of the replacement fi lter, taking note to align the fl ow arrow of the fi lter with 
the fl ow arrow of the instrument top panel.  CAREFULLY slide the tubing back into the internal portion of the 
instrument and seat the new fi lter into the retaining clip.  Inspect the sample hoses for signs of oil entrapment.  
Replacement of the sample fi lter usually requires cleaning or replacement of the sample hoses.

5.4  Replacing the Sample Hoses

Inspect the inside diameter of the tube for signs of oil build up, dirt, obstructions, kinks, cuts, fraying, or any 
other signs of wear.  Oil contamination cannot be cleaned out of sample hoses due to the density of the oil 
restrictor.  If oil is visible in the Low Pressure Vapor hose, replace the hose and restrictor with P/N 2100-0010.

For the High Side Liquid Sampling Hose, inspect the inside diameter of the tube for signs of oil build up, dirt, 
obstructions, kinks, cuts, fraying, or any other signs of wear.  Oil contamination cannot be cleaned out of sample 
hoses due to the density of the oil restrictor.  If oil is visible in the High  Pressure Liquid Trap Hose, replace the 
hose and restrictor with P/N 2100-0010.

To replace the Restrictor Assembly, do the following:

1. 

Disconnect the sample hose from the analyzer

2. 

Remove the brass fi lter (with hose attached) from the coupler and discard.  Be sure to use a 
backing wrench so as not to damage the coupler.

3. 

Check for signs of oil and debris in the coupler.

4. 

Using a cleaner that contains ONLY Tetrachloroethylene and carbon dioxide, spray all parts of 
the coupler with the cleaner to remove the oil, per the cleaner’s instructions.  DO NOT soak the 
park for more than 60 seconds.

5. 

Allow couple parts to dry. Check coupler parts for oil again. Failure to clean the oil out of the 
coupler will result in premature clogging of the new fi lter.

6. 

Install the new fi lter (P/N 2100-0010), into the coupler and lightly tighten.  Finger tight is        
suffi cient.

5.5  Changing the Printer Paper

Pur•Chek™ Pro Diagnostic Refrigerant Analyzers use a custom inexpensive thermal paper for printing.  The 
paper roll should be changed when a red stripe appears on the left side of the printout.

To change the paper roll, press the “HELP” button on the “Ready to Air Cal” screen as shown in Section 4.2, 
Figure 3.  The screen will display several options as shown in Figure 22.  Press the “SET” button to advance to 
the screen shown in Figure 23 and then press the “FEED” button.

Содержание PUR-CHECK PRO

Страница 1: ...Product Leadership Training Service Reliability Instruction 2100 9001 Operation and Maintenance Rev 0 June 2010...

Страница 2: ...LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II ANY OBLI GATION LIABILITY...

Страница 3: ...3 6 Back Panel Connections 7 3 7 Hard Shell Storage Carrying Case 7 4 Operation 8 4 1 Battery Installation 8 4 2 Turning On the Unit 9 4 3 Calibration 9 4 4 Vapor Sampling 10 4 5 Liquid Sampling 10 4...

Страница 4: ...Pur Chek Pro Instruction 2100 9001 ii...

Страница 5: ...rformed or adhered to could result in minor or moderate injury IMPORTANT The use of the word IMPORTANT in this manual calls attention to a procedure practice condition or the like which if not correct...

Страница 6: ...tain or will not emit heavy loads of oil or liquid NEVER admit any sample into the instrument at pressures in excess of 500 psig NEVER obstruct the air intake sample exhaust or case vent ports of the...

Страница 7: ...The instru ment utilizes non dispersive infrared NDIR technology to determine the weight concentrations of multiple refrigerant types Refrigerant purity is displayed on the LCD Screen The user must de...

Страница 8: ...rared NDIR Refrigerant Sample Size 0 3 ounces 8 5 grams per sample Power External Power Supply 110 220V 5 60Hz 0 6A External Battery Charger 110 220V 50 60Hz 1 95A DC Supply Requirements 12 15 VDC 2 A...

Страница 9: ...void the warranty Printer Door Air Intake Port Sample Filter Control Panel 3 2 Low Side Vapor Sample Hose The 6 foot 1 8 meter Low Side Vapor Sample Hose is constructed of a polyurethane tube A brass...

Страница 10: ...AE Flare Nut 3 4 AC Power Adapter The Pur Chek Pro is powered via a 110 220 VAC 50 60 Hz power transformer This transformer is included with each unit and converts a standard 110 220 VAC 50 60 Hz wall...

Страница 11: ...let port free and clear at all times Do not operate near open flame Sample Inlet Sample Outlet Battery Charge Port 12 VDC Power Input AC Adapter 3 7 Hard Shell Storage Carrying Case The hard shell sto...

Страница 12: ...rant Type R 12 R 134a R Miscellaneous Refrigerant Equipment 12 VDC 1A Design Pressure 300 psig CFC 12 R 12 or HFC 134a R 134a R 134a For Accuracy see Manual detect 2 or greater of air in HFC R134a to...

Страница 13: ...alibration takes 95 seconds Figure 4 and brings fresh air into the unit via an internal pump This fresh air purges any excess refrigerant from the unit and ensures accurate test results Calibration re...

Страница 14: ...om pletely depressed 3 Connect the appropriate end of the hose to the instrument and the opposite end of the liquid sampling assembly to the High Side Liquid port of the system or cylinder Affix the o...

Страница 15: ...lay the detail screen shown in Figure 11 If the blend or refrigerant mixture is not recognized the screen shown in Figure 12 will be displayed BACK MORE Figure 10 Figure 12 Figure 11 RESULTS R410A CON...

Страница 16: ...procedure To exit the test press the EXIT button Figures 15 16 17 show sample printouts for various test results NOTE Care must be taken to leave a clean edge when tearing off the printed results Neu...

Страница 17: ...he ELEV button will take you to the elevation screen shown in Figure 19 Use the UP DOWN buttons to adjust the unit to the elevation in your area Press the SAVE button when finished to return to the he...

Страница 18: ...Oil contamination cannot be cleaned out of sample hoses due to the density of the oil restrictor If oil is visible in the Low Pressure Vapor hose replace the hose and restrictor with P N 2100 0010 Fo...

Страница 19: ...lled properly Slide the paper through the slot in the printer door and close the door 5 6 Error Messages In the unlikely event that an Error message is displayed on the screen power off the unit take...

Страница 20: ...2100 0010 Spare Low Pressure Sample Hose Filter 2100 0012 Spare Rechargeable Battery 2100 9001 Instruction Manual 7 Service Centers Service and replacement parts can be found by contacting any of the...

Страница 21: ...Pur Chek Pro Instruction 2100 9001 17 NOTES...

Страница 22: ...ers 621 Hunt Valley Circle New Kensington PA 15068 PH 724 334 5000 Toll Free 800 736 4666 Fax 724 334 5001 Website www mybacharach com Email help mybacharach com Printed in USA Registered Trademark of...

Отзывы: