Instruction 24-9351
PCA
5.9 Calibrate CO
(For PCA Models 20, 25, 30, 35, 50, 55, 60 & 65)
Material Required:
• Calibration Gas Fixture (Section 5.2)
• Gas Cylinder, 500 ppm CO in air (Refer to Section 8.2)
• Gas Cylinder, CO (1000 ppm) and H
2
(1000 ppm) in Nitrogen
(Refer to Section 8.2)
Procedure:
1. Enter the Calibrate Menu Screen per Section 5.3. Then choose to
calibrate the CO sensor.
2. At the conclusion of Step 1 the pump should start running.
3. Using the Gas Fixture shown in Figure 5-1, attach the CO calibration-
gas cylinder to the analyzer’s GAS inlet.
4. Adjust the regulator of the calibration fixture for a flowmeter indica-
tion of approximately 2 SCFH.
5. Wait until the MEASURED reading on the screen stabilizes (approxi-
mately 3 minutes). Then use the
st
and
keys to enter an
APPLIED value* that equals the concentration which is stamped on
the CO calibration-gas cylinder.
Typical CALIBRATE CO Screen During Calibration Procedure,
Using 500 ppm CO Calibration Gas:
CALIBRATE CO
MEASURED 492 PPM
APPLIED 0500 PPM
6. Press the
key to calibrate the analyzer’s MEASURED value to
that of the APPLIED value. At this time the Test Gas CO/H2 Screen
is displayed.
* The calibration range for this screen is 250 – 1500 ppm. Any attempt
to calibrate outside this range will cause the analyzer to display an
error message.
5-9
Calibration
Содержание Portable Combustion Analyzer 24-9351
Страница 48: ...PCA Instruction 24 9351 4 32 NOTES Operation ...
Страница 60: ...PCA Instruction 24 9351 Calibration 5 12 NOTES ...
Страница 81: ...Instruction 24 9351 PCA 8 3 Figure 8 1 Replacement Parts Parts Service 2 1 3 ...
Страница 83: ...Instruction 24 9351 PCA Figure 8 3 Replacement Parts 8 5 Parts Service 2 0 1 9 1 8 2 5 ...
Страница 94: ...PCA Instruction 24 9351 A 10 Appendix A NOTES ...
Страница 99: ...Instruction 24 9351 PCA ...