Microprocessor Module Replacement Kit for H25-IR Infrared Gas Leak Detectors 

 

 

3015-5575 Rev 0

 

 

7.  Reconfiguring the H25-IR (Replacement of Pre Version 1.4 Modules Only) 

 

 

NOTE

:  Customized system settings are automatically maintained on units that had newer firmware (version 

1.4 and newer).  However, these settings are reset to their factory default values when replacing the 

microprocessor board on units with older firmware (before version 1.4).  This section ONLY applies if you 

started with an H25-IR having a firmware version prior to 1.4. 

 

 

 

IMPORTANT

:  A  reset to factory defaults occurs  automatically on the first boot up with the new module.  

Afterwards, the factory menu will be enabled to facilitate the reconfiguring of previous system settings that 

were lost as part of the pre 1.4 microprocessor upgrade. 

 

 

IMPORTANT

:  Prior to adjusting settings in the Factory menu, you MUST first adjust the Emitter Power and 

DigiPot settings (in that order) in the  Diagnostics menu.  Then adjust leak source information, followed by the 

optional reconfiguration of any other settings that you may have recorded earlier. 

 

 

NOTE

:  Refer to the H25-IR Instruction Manual originally shipped with your unit for additional detailed 

information  about  configuring the IR bench’s emitter power, DigiPot, and leak source settings.  If you no 

longer have the manual that shipped with your device, the standard H25-IR manual may be downloaded from 

the Bacharach website at 

www.mybacharach.com

.  If you have any questions, you may contact Bacharach’s 

Customer Service Department.

 

 

 

1.  Adjust the IR bench emitter power first.  Refer to the H25-IR manual for details. 

2.  Adjust the IR bench DigiPot setting.    Refer to the H25-IR manual for details. 

 

NOTE

:  Following the reset  after the microprocessor replacement,  the factory menu will 

automatically be enabled.  Access the factory menu and perform the following steps.

 

 

 

NOTE

:  For H25-IR models with an internal leak source, the leak source  parameters for the 

installed leak source will need to be entered following the reset. These parameters (leak rate and 

leak temp) are printed on the ID tag attached to the leak.

 

 
 
 

3.  Adjust the LEAK RATE value to match the internal leak rate in grams per year.  This information can be found 

on the tag attached to the internal leak source. 

4.  Adjust the LEAK TEMP value to match the internal leak temperature in degrees C. 
5.  Restore any additional setup and/or system settings as needed.  Refer to the appropriate H25-IR manual for 

details. 

6.  At this point the reconfiguration is complete. 

 

 

 

 

  

  

 

 

 

 

 

 

 

 

 

 

Содержание Microprocessor Module Replacement Kit

Страница 1: ...ming firmware with a simple plug in microprocessor module minimizing downtime This instruction manual describes how to replace or upgrade the microprocessor module in your H25 IR infrared gas leak det...

Страница 2: ...ocess is straight forward additional steps may be required depending on the version of H25 IR firmware that is currently installed Firmware version 1 4 is the point after which system settings and con...

Страница 3: ...e flowchart before beginning the replacement process Most steps in this microprocessor board replacement procedure are explained in this document however you may need to refer to the H25 IR Instructio...

Страница 4: ...ing so you can reenter them after you re done Users having pre version 1 4 firmware must adjust IR bench emitter DigiPot and leak source settings For details on these processes refer to your H25 IR ma...

Страница 5: ...or H25 IR Infrared Gas Leak Detectors 3015 5575 Rev 0 5 Figure 3 Location of Microprocessor Module Inside the H25 IR Housing Figure 4 Orientation of Microprocessor Module Microprocessor Module Note Po...

Страница 6: ...ard by the edges only 2 Orient the new microprocessor so the gold tab and hole are at the bottom left corner refer to Figure 4 3 Align the two rows of pins on the back of the microprocessor board with...

Страница 7: ...hipped with your unit for additional detailed information about configuring the IR bench s emitter power DigiPot and leak source settings If you no longer have the manual that shipped with your device...

Страница 8: ...Detectors 8 3015 5575 Rev 0 Headquarters 621 Hunt Valley Circle New Kensington PA 15068 7074 Toll Free 1 800 736 4666 Tel 1 724 334 5000 Fax 1 724 334 5001 Website www MyBacharach com E mail help MyBa...

Отзывы: