Manual Part Number:
5-06-7000-81-1, 7-
08-1234-40-0
Manual File:
069706 Rev. A
Page 1-12
Liquid Sample Hose
The 6.5-foot (2 meter) Liquid Sample Hose configuration is constructed of a polyurethane tube with
an oil reservoir. A Brass Sample Hose Restrictor acts to transform Liquid refrigerant to Vapor at the
sample connection point while the Liquid Sample Trap assembly collects oil and provides a means
of expulsion after the test is complete. The syringe is provided with a magnet for attaching it to the
tank, an instrument inlet port mating connector on one end and a ¼” SAE female flare coupling nut
on the service end. The maximum pressure is 500 psig (35 Bar).
NOTE:
The analyzer will indicate “Error #5” if the analyzer does not receive a good sample due to
obstructed flow or lack of flow. If this occurs the Brass Sample Hose Restrictor may need to be
replaced.
AC Power Adapter
The
Legend Series
™ is powered via a rechargeable Lithium-Ion battery. You can also power the
unit with the AC Power Adapter which converts a standard 100-240VAC 50/60Hz wall outlet to
12VDC, 1.6A. This AC Power Adapter will charge the battery when connected to the analyzer.
NOTE:
Use of any other power source may cause damage to the unit and void the warranty.
Service End
(1/4 SAE Flare Nut)
Analyzer End
Liquid Trap Assembly
Brass Sample Hose
Restrictor