3.  If Pressure, Vacuum, are equal to Ambient, 14.xx to (12.xx if you are in Denver). 

Then there are three possibilities; 

a)  The pump is not pumping.  

i.  Note

the pump works on a volume / differential pressure and cannot 

be verified by putting your finger over the end to see if it’s working. 

b) 

There is a hose off of a connection

 

c) 

The Manifolds are stuck open; this could be a result of dirt inside the manifold 
holding a valve open mechanically. It could also be a result of a failed manifold 
control board, holding the solenoids on the manifolds open electrically. (see next 
section on troubleshooting a Manifold control board failure)

 

 

Troubleshooting a suspected Manifold control board failure 

If the fault code is an 800 or 1800 it could be due to a manifold control board failure. 
After checking to make sure the pump is functioning properly the follow procedure 
should be used to check the control board. 

The symptoms are: during normal sampling mode the Pressure value and the ambient 
value are equal or very close to equal at sea level they would be 14.xx PSIA. The 
pressure is the live number with the Vacuum and Ambient values being stored for 
reference during a purge and pressure check cycle. A good way to verify that the 
manifold control board has failed electrically and is holding 2 or more solenoids open is 

power down the monitor first,

 and then disconnect (unplug) all the solenoids from the 

board. Then power the monitor up and If the Pressure goes to maximum Vacuum 
(somewhere between 10.8 and 11.2) with all the manifolds disconnected then you know 
it’s an electrical failure of the manifold control board. If the values stay equal at ambient 
pressures then you know it’s a mechanical failure from dirt or moisture holding the 
valves open. 

Do not reconnect the manifolds with the power on. 

 

REZERO VOLT TOL (4000) –

 

 

1. 

The detector output voltage is out of tolerance. 

(see HGMMZ voltage adjustment 

in the annual maintenance guide or the HGMMZ voltage adjustment bulletin) 

 

OVER RANGE DETECTED (2000) – 

 

1.  This indicates the measured concentration on one or more zones exceeds the 

measurement range.  

2.  Has the monitor been exposed to pure refrigerant? If not check the sensor 

voltage range. 

 

Содержание HGM-MZ

Страница 1: ...ASURABLE DIFFERENCE New Kensington PA USA Dublin Ireland Toronto Canada mybacharach com BACHARACH Inc HGM MZ Multi Zone Monitor Annual Maintenance And Troubleshooting Guide Service Testing and Mainten...

Страница 2: ...f the monitor is installed in areas of high levels of dust and particulate it is recommended that semi annual or quarterly filter changes be made to prevent system faults due to purge or zone restrict...

Страница 3: ...screen on the HGM MZ select the top left block under STATUS and press enter This will take you to the SETUP screen To enter service mode From the main sampling screen figure 1 press enter two times th...

Страница 4: ...voltage and IR voltage adjustment screen select SERVICE MODE ENTRY press enter twice then select SYSTEM and press enter To enter service mode From the main sampling screen figure 1 press enter two tim...

Страница 5: ...e the arrow keys move to it then press enter which will allow you to see the voltage and make the adjustments with the up and down arrow keys The detector digipot DET DIGIPOT voltage range should be b...

Страница 6: ...ot set outside the range of 300 to 500 after adjusting the IR detector voltage IR DIGIPOT A typical MW setting for a sensor with Yellow tape is 325 without the tape it is approx 425 To adjust the MW u...

Страница 7: ...The pinch test can isolate the problem quickly In the diagnostic screen observe the live pressure PRES then pinch the tube on one side of the white disc Hydrophobic filter pinching either side is OK...

Страница 8: ...ry 0204 0020 Infrared Bench Replacement Kit 3015 4572 Power Supply Board 3015 5149 Power Entry Board w Fuse Manifold Control Board 3015 5095 Main Processor Board Motherboar d Replaceme nt Fuses 2 0004...

Страница 9: ...A Dublin Ireland Toronto Canada mybacharach com Inside Cover View Filters 3015 3125 Charcoal Filter Assembly with Fittings 3015 5931 Display PC Board and Keypad PC Board Kit With interconnecting Ribbo...

Страница 10: ...ASURABLE DIFFERENCE New Kensington PA USA Dublin Ireland Toronto Canada mybacharach com Figure 1 07 1650 Hydrophobic Filter Figure 2 3015 2906 Line end filter only Figure 3 3015 3420 Line End Filter A...

Страница 11: ...3 6 Gas test zone with bump gas 7 Check IR voltages both Detector and Digipot 8 Check sample pump pressures 9 Check HGM RD remote display If applicable 10 Take monitor in and out of service mode to ca...

Страница 12: ...000 REZERO VOLT TOL and a 200 GAIN SET FAULT combined to indicate two faults registered at the same time Most common fault code is an 1800 which is a combination of a 1000 No flow on purge and an 800...

Страница 13: ...log If filter is clear then the problem could be in the manifolds or manifold control board Call Bacharach tech support for assistance in trouble shooting the Manifold control board 724 334 5000 DET I...

Страница 14: ...mode the Pressure value and the ambient value are equal or very close to equal at sea level they would be 14 xx PSIA The pressure is the live number with the Vacuum and Ambient values being stored fo...

Страница 15: ...the dual 4 20 ma dc option card was installed and one or both current loops are open unconnected to load or without shorting jumpers 2 Check the wiring to load monitoring circuit on both 4 20 mA loops...

Страница 16: ...nd burst of gas directed at the test zone pickup while the monitor is polling that zone f After a short delay the monitor will indicate the presence of gas on that zone The PPM reading will vary as th...

Страница 17: ...Bacharach THE MEASURABLE DIFFERENCE New Kensington PA USA Dublin Ireland Toronto Canada mybacharach com...

Отзывы: