Make sure that the filter is firmly
seated against the fine screen.
4. Replace the rubber probe tip.
Replacing Airflow Ball Indicator:
1. Remove the clear plastic section of
the probe tip by gently pulling and
twisting the white grip.
2. Turn the probe tip upside-down and
tap on it to remove the old airflow
indicator ball.
3. Insert the new airflow indicator bal
into the tip.
Filter
Rubber
Probe Tip
Airflow
Indicator Ball
Probe Tip
Caution: Equipment Damage
Hazard. Submerging the probe in
liquid will damage the vacuum pump.
Exposing the probe to a stream of
refrigerant coming out of a tank will
severely reduce the life or destroy the
sensor. The life of this unique rare
earth sensor is directly proportional to
the amount of refrigerant that passes
through it. With normal intermittent
service, the sensor can last for one
year or longer. Exposure to high con-
centrations of refrigerant may require
adjustment of sensor heat. To check,
repeat the test procedure in the
“Heater Adjustment” section.
4. Reattach the probe tip to the probe
assembly.