Bacharach GDX-150 Скачать руководство пользователя страница 2

GDX-150 Sensor Transmitter 

 

3.  Specifications 

 

Category 

Description 

Product Type  

GDX-150 transmitter for low level continuous monitoring of ammonia in 
ambient environments 

Coverage  

Single sensor, 40-foot diameter 

Detector Type  

Electro-chemical, field replaceable 

Sensitivity  

1 ppm  

Measuring Range  

0-500 ppm  

Accuracy  

±15% of full scale, ±1 count (whichever is greater) of reading from 0-500 ppm  

Long term Drift  

<10% per 6 months  

Linearity  

<10% of full scale  

Temperature Drift  

<0.1% per degree C over temperature range  

Housing  

316 stainless, suitable for Class 1, Div 1 and 2, Gr. B, C, D  
¾ inch NPT threading  

Calibration  

Through GDA-400 or GDA-1600 controllers  

Response Time  

T

50

 <30 s , T

90

 <90s  

4-20 output Signal  

2-wire, 4-20 mA current sink. Max output = 25 mA  

Operating Temp  

-20 – 50 °C (-4 - 122 °F)  

Ambient Humidity  

5% to 90% RH (non-condensing)  

Power  

9 – 36 VDC, 250 mA (@ 24 VDC)  

Certification  

Intrinsically safe with use of an IS barrier  

Warranty  

2 years from date of shipment, consumables not included 

 

4.  Accessories and Replacement Parts 

 

Part Number 

Description 

5600-0004

 

Calibration Cup/Sample Draw Adaptor

 

5600-0003

 

Splash Guard/Remote Cal-Cup

 

0024-7059

 

Calibration Kit – Regulator, Tubing and Carrying Case (no gas included)

 

0051-3160

 

NH

3

 Calibration Gas, 25 ppm, 58 Liter

 

0051-7131

 

Zero Grade Air, 103 Liter

 

5600-5003

 

NH

3

 Replacement Sensor

 

 
 
 
 

 

 

Headquarters: 

621 Hunt Valley Circle, New Kensington, PA  15068-7074 

Toll Free: 800-736-4666 • Tel: +1-724-334-5000 • FAX: +1-724-334-5001 

Website: www.MyBacharach.com • E-mail: [email protected] 

 

2  

P/N: 5600-9000 Rev 2

 

Отзывы: