Instruction 0024-9396
17
Fyrite Pro Quick Reference Guide
Figure 3. Printer Alignment & Sample Printout
B A C H A R A C H , I N C .
F Y R I T E P R O A N A L Y Z E R
= = = = = = = = = = = = = = = = = = = = = = =
D A T E : 2 9 / 1 0 / 2 0 0 3
T I M E : 1 4 : 4 5
F U E L : ( F 1 )
q A 1 5 . 9 %
E T a 8 4 . 1 %
E f f i c i e n c y 7 4 . 5 %
L a m b d a 1 . 4 1 %
T A 3 2 9 ° C
T L 3 0 . 7 ° C
O 2 6 . 1 %
C O 2 9 . 3 %
C O 5 7 p p m
C O U n d i l u t e 8 0 p p m
C O / C O 2 . 0 0 0 6
D r a f t 0 . 0 0 m B
Notes:
When a calcuation cannot be
made because of improper data (i.e.,
oxygen level above 17.9%), four dashes
“- - - -” appear in place of the calculated
value on both the screen and printout.
When one or more of the following
sensors are not installed, their associated
values do not print:
Oxygen Sensor:
Fuel, Efficiency, Excess Air, Stack
Temp, Primary Temp, O2, CO2, and
CO Air Free
CO Sensor:
CO, and CO Air Free
Draft Sensor:
Draft
Either Thermocouple:
Delta
Temp