CO Appliance Kit 

 

 

0024-9498 Rev 1 

Step 

Description 

Check the probe’s filter for signs of soot and/or moisture.  Replace if 

necessary. 

Attach the desired probe to the probe handle.  Attach the probe tubing to 

the GAS inlet on the bottom of the instrument.  Use the female gas 

connector to attach to the INSIGHT

 Plus analyzer. 

 

 

NOTE:  

The first screen will show a “T-STK” error as the 

thermocouple is not used in this test.  Press the “ENTER” button 

to step past this screen.

 

      

Power on the analyzer and wait while it initializes.  Follow the instructions 

provided in your analyzer’s instruction manual. 

If using either the InTech® or INSIGHT® Plus combustion  analyzer,  confirm 

that your  analyzer is configured for Siegert combustion equations.  (Go to 

the COMBUSTION EQUATIONS option in the SETUP MENU for more 

information.) 

When instructed to perform a CO test based on EN7967 procedures, place 

the CO appliance probe as instructed in EN7967. 

Navigate to the CO AMBIENT TESTING option of the analyzer’s MAIN MENU. 

Follow the on-screen instructions to begin the CO testing. 

10 

After the testing is complete, analysis results will be displayed. 

11 

Review the results based on the criteria set forth in EN7967. 

12 

Make appliance and/or installation corrections, then re-test as required. 

 

  

  

 

 

 

 

 

 

 