Operation of Baby Control Digital breathing monitor –
TWIN MODE (BC-220i, BC-230i)
The numerical display (2) will show the intensity of breathing movements detected by Baby Control Digital approximately eve-
ry 2.5 seconds („9“ - the most intensive, „1“ - the least intensive, „0“ - no movement detected). At first, the value coming from
the left socket (9) is displayed – breathing movements of the baby lying on the sensor pad to the left. Then the value coming
from the right socket (10) is displayed - breathing movements of the baby lying on the sensor pad to the right. At the same
time the indicator light (6) above the socket (9) and the indicator light (7) above the socket (10) flash green if the intensity of
detected movements reaches or exceeds the pre-set threshold level. Thus it is possible to check even from a distance whether
the device has detected movements that are intensive enough. An acoustic warning signal will sound from the speaker (3)
and the indicator light (6 or 7) will start to flash red when the measured value does not reach the pre-set threshold level of
detection four times in succession (12 seconds). Then the alarm is activated. If the measured value is very low or equal to
zero, it is likely that the mattress placed on the sensor pad is extremely thick or baby is lying close to the edge of the sensor
pad (11). Please change this state to avoid false alarms.
IMPORTANT:
In the twin mode, at first the status of the sensor pad connected to the socket on the
left-hand
side
flashes on the display, then the status of the sensor pad on the
right-hand side
.
IMPORTANT:
For the Baby Control Digital breathing monitor operation it is very important to adjust the
threshold level of detection, i.e. the value between „1“ and „9“ to be attained by the breathing movement
of the baby. Typically, the value „3“ is used (factory setting). You may change the setting in justified cases (see
Adjusting the threshold level of detection).
Alarm status of Baby Control Digital breathing monitor –
TWIN MODE (BC-220i, BC-230i)
1. The detected value of breathing movements coming from the left or right sensor pad (11) is lower four times in succession
(12 seconds) than the pre-set threshold level of detection.
2. The indicator light (6 – baby lying to the left or 7 – baby lying to the right) above the appropriate socket starts flashing red
and simultaneously four warning acoustic signals sound from the speaker (3). (If detection of breathing movements is
restored in this phase, Baby Control Digital continues its normal operation.)
3. The alarm is triggered! The display (2) shows the letters „H” - „E” - „L” – „P”.
4. The alarm sounds until you press the on/off button (1). Then the device switches off.
5. The alarm is not silenced even when the signal from any of sensor pads (11) is transmitted again.
6. In case of any alarm you must check whether it is not false (false alarm can be caused by a mattress being too thick or baby
moving outside the sensor pad).
IMPORTANT:
When you use more than one sensor pad under one baby in the twin mode, you may connect
maximally 2-2 sensor pads to Baby Control Digital breathing monitor using the T-connection (part of delivery).
Take special care that the two sensor pads connected using T-connection are placed under the same baby
!
Otherwise the device is not able to differentiate between the movements of two babies.