BABBITT LS7500 Скачать руководство пользователя страница 10

www.babbittinternational.com

 

Reference Manual 

 

2021

 

 

10 

 

3.4 Physical Installation 

 

1)   Install the LS7500 Presence/Absence into a recommended or “correct” section of pipe as outlined in          
 

Figures 3-1 to 3-6

.  

2)  Position the orientation of the flange sensor as noted – refer to 

Section 3.2.1

 

Partial-ring 

or 

Section 

 

3.2.2 Full-ring. 

 

3)  Mount the sensor between two 150# or 300# flat faced flanges (raised faced only if ordered raised 
 

face). For use with other flanges contact factory. 

4)   Use full face gaskets compatible with the process material – refer to 

Section 3.2.3.

 

5)  Tighten the flange bolts evenly to insure a proper seal.  
 

 

a) All bolts should be put in place and evenly hand tightened. 

 

 

b) All bolts should be tightened in 10 ft-lb increments using a star pattern. 

 

 

c) Torque all bolts to 50ft-lb max. 

6)   The LS7500 has a 1” NPT conduit entry. When wiring these units, conform to the National Electrical 
 

Code and any other city or company codes. 

       7)   Always install the electrical connection into the 1” NPT on the LS7500 so water will not follow the 

connection routing into the threads of the 1” NPT hub. To guarantee that condensation and water will                  

 

stay outside the unit, install an EYSM unilet before each LS7500 and fill with explosion proof sealing 

 

cement. 

 

       

CAUTION: 

Always take the necessary safety precautions when cutting or welding. 

a) Tag and lock out the electrical power to the equipment that services the vessels. 
b) Check liquid or dry powders that create a gas in the vessel for oxygen as well as the  

 

     explosion factor. (All dust will explode.) 
c) Vessels that are cross vented have to be isolated. 

        

 

3.5 Grounding  

 

 

      Proper installation of the LS7500 is critical to help ensure reliable operations. The most important  

 

      installation consideration is a proper sensor ground return. If the vessel is non-metallic (i.e. plastic or       

 

      plastic lined), a separate ground return must be provided. In such case, the ground return must consist of a       

 

      separate metal mounting flange, installed into a pipeline, adjacent to the sensor flange. A ground wire must 

 

      be installed between the ground screw on the sensor flange and the ground screw on the metal ground     

 

      flange, as shown in 

Figure 3-7. 

 

3.6 Ground Continuity Test 

 

 

 

       

Refer to 

Figure 3-7

 below. 

 

       1)   Turn power to LS7500 off. 

 

       2)  Use ohm meter on lowest scale to check following: 

 

 

 

a) Point A to B – less than 1 ohm 

 

 

 

b) Point B to C (both ends of metal pipeline) – less than 1 ohm 

 

 

 

c) Plastic Pipeline ONLY – Point B to D – less than 1 ohm 

 

Содержание LS7500

Страница 1: ...LS7500 Presence Absence Switch OWNERS MANUAL INSTALLATION CALIBRATION TROUBLESHOOTING WARRANTY P O Box 70094 Houston TX 77270 713 467 4438...

Страница 2: ...arise during installation please contact Customer Service 1 800 835 8012 or 713 467 4438 8 00 a m to 5 00 p m CST US Failure to follow safe installation and servicing guidelines could result in death...

Страница 3: ...ring Flange Location 7 3 3 2 Full ring Flange Location 8 3 3 3 Full faced Gaskets Flanges 8 3 4 Physical Installation 10 3 5 Grounding 10 3 6 Ground Continuity Test 10 3 7 Removing the Electronics 11...

Страница 4: ...ate when material is detected 1 2 Specifications Electrical Power 115 VAC 15 50 60 Hz 2 watts standard 12 VDC 24 VDC or 230 VAC optional Output 2 Form C contacts DPDT relay 5 amp resistive At 125 250...

Страница 5: ...al com Reference Manual 2021 5 1 3 Ordering Information LS7500 115VAC FP 4 150 PR Model LS7500 Power Supply 115VAC 24VDC 12VDC or 220VAC Sensor Type FP Flange Probe Sensor Size Flange Model PR Partial...

Страница 6: ...www babbittinternational com Reference Manual 2021 6 1 4 GA Drawing Figure 1 1 1 5 Sensor Rings Figure 1 2...

Страница 7: ...ting voltage of the LS7500 is marked on the power supply board Hold the unit so that nothing is touching the flange sensor element is and press the test switch When the test switch is depressed the re...

Страница 8: ...run when they should have been turned off 3 3 2 Full ring Flange Location Full ring sensor flanges are primarily used for non conductive media such as oils and hydrocarbons These sensor flanges can be...

Страница 9: ...www babbittinternational com Reference Manual 2021 9 SENSOR FLANGE MOUNTING...

Страница 10: ...n and water will stay outside the unit install an EYSM unilet before each LS7500 and fill with explosion proof sealing cement CAUTION Always take the necessary safety precautions when cutting or weldi...

Страница 11: ...ower supply card When reinstalling this card is made so that it cannot be plugged in backwards 2 To remove the power supply card you must a Disconnect supply power at main power source b Remove sensin...

Страница 12: ...ER SUPPLY IS INTENDED FOR ONLY ONE SUPPLY VOLTAGE THIS VOLTAGE IS PRINTED ON THE POWER SUPPLY BOARD FOR 115 VAC OR 230 VAC UNITS HOOK UP THE HOT LEAD TO L1 AND THE NEUTRAL TO L2 WITH APPROPRIATE GROUN...

Страница 13: ...www babbittinternational com Reference Manual 2021 13...

Страница 14: ...www babbittinternational com Reference Manual 2021 14...

Страница 15: ...www babbittinternational com Reference Manual 2021 15...

Страница 16: ...babbittinternational com Reference Manual 2021 16 4 CALIBRATION 4 1 Sensitivity Calibration Conductive Materials NOTE The green LED indicates relay status The relay is energized when the green LED is...

Страница 17: ...www babbittinternational com Reference Manual 2021 17 4 2 Sensitivity Calibration Non Conductive Materials NOTE The green LED indicates relay status The relay is energized when the green LED is on...

Страница 18: ...timing ranges you can set the timer for a shorter range then change the range Switches S1 and S2 to increase the time For example the desired time delay is 15 minutes This is equal to 900 seconds By...

Страница 19: ...No power to unit Check for correct power to unit Blown on board fuse Replace fuse Bad sensing card Repair or replace card RED LIGHT remains on at Sensing element is shorted to Unplug banana plug and...

Страница 20: ...rding to instructions Babbitt International Inc will repair or replace at its option FOB at its plant or any other location designated any part which proves to be defective in manufacture or workmansh...