7
Key symbol
Normal weighing state
Set state
Weight maintenance key
Short press
→
F2.1 = 1, keep/cancel.
F2.1 = 2, switch between percentage and weight.
F2.1 = 5, switch between quantity and weight.
Long key
→
enter setting menu.
Return to the last menu.
Accumulation key
Short key
→
F2.1 = 4, include display weight in accu-
mulation value.
Long key
→
F2.1 = 3, select scale to sample target
weight.
F2.1 = 4, accumulate weight of scale display.
F2.1 = 5, count sampling of scale.
No definition.
Unit conversion key
Short key
→
in weighing state, switch weight unit. The
corresponding unit indication lamp is on.
Flicker bit is on the left.
Skin removal key
Short key
→
net weight turns to gross weight; induction
lamp of net weight “Net” is off.
Flicker bit is on the right.
Skin removal key
Short key
→
gross weight turns to net weight. Indication
lamp of net weight “Net” is on. Conduct skin removal
operation for multiple times.
Digit flicker position reduces.
Clearing key
Gross weight state resets weight. When the scale is in
net weight, dynamic state, saving state and out of reset-
ting range, clearing operation is invalid
In setting, digit of flicker posi-
tion increases. In adjustment
of display, accumulation is
cleared.
ON/OFF key
Short key
→
start up or print. (only for printer models)
Printing format refers to Appendix1.
Long key
→
shut down. Turn on/off
Confirm operation, to save
setting data.
V
INTRODUCTION
TO
PANEL
SETTING
ENTRY:
Press the
button on the operating panel in the state of normal weighing.
If F1.14 = 0, you can set all the parameters within F1~F5.
If F1.14 = 1, you can only set all parameters within F2~F5.
If F1.14 = 1 and you need to set the parameters within F1 menu, you can press the calibration switch button until the F1 menu is
entered.
USER’S MANUAL
TRANSERENITE ARX