Nebulizer elements cleaning and maintenance
1. Remove the jet nebulizer from the nebulizer and dismantle each part by
following steps from 1 to 4 described in “Use Instructions”.
2. Once the three parts are dismantled (B) submerge each part into a solution of
three parts of warm water and one part of white vinegar for 30 minutes, or use a
bacteria germicidal disinfectant available in drug stores.
3. Drain any liquid from components, reassemble and insert them into the unit.
4. Turn on the device and leave it on for about 5 minutes. This operation will dry
components and avoid any medicine wastage, which may obstruct air flow or any
new medicine.
Warning:
nebulization elements are to be replaced about 12 months after
use, or if the unit has not been used for a long time. Use only the manufacturer
original spare parts (B.Well).
8. Air Filter Change
1. The filter should be replaced if it gets dirty or worn out.
2. Remove the jet nebulizer from the nebulizer body.
3. Turn the nebulizer base upside down.
4. Remove the filter cover lid and replace it with a new one.
5. Remount by pressing the filter cover lid, with the new filter in place.
Precaution:
using a dirty filter or replacing it with another material such as
cotton, may damage the compressor.
11