
24
Instructions for use
En
WF-1000
Recommended separation distances between
portable and mobile RF communications equipment
and the WF-1000 Thermometer
The WF-1000 Thermometer is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the
WF-1000 Thermometer can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmit-
ters) and the WF-1000 Thermometer as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum
output power of
transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d = 1.2 P
80 MHz to 800 MHz
d = 1.2 P
800 MHz to 2.5 GHz
d = 2.3 P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmit-
ter in watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and refl ection from structures, objects and people.
Содержание WF-1000
Страница 1: ...Instructions for use EN WF 1000 Infrared Ear Forehead Thermometer...
Страница 2: ......
Страница 27: ......
Страница 28: ...IM_WF_1000_EN_0320 www bwell swiss ch B Well Swiss AG Bahnhofstrasse 24 9443 Widnau Switzerland...