background image

4

GENERAL INFORMATION

•     The TCW2043SDD consists of 2 separate units : a refrigerator and a freezer.
•     before using the TCW2043SDD, read these operating instructions carefully, including all the

information on operating safety, use and maintenance.

•     keep these operating instructions ready at hand and leave them with the unit, so that all users can

find out about the functions and safety regulations.

•     Ensure that there is sufficient room around the unit for air circulation. Please refer to the

installation information regarding this point.

•     The refrigerator as well as the freezer compartment has a bank of    waterpacks filled with water and

distributed inside the cabinet. During operation the waterpacks freeze and act as cold storage for

protection of the vaccines as well as the icepacks stored in the TCW2043SDD during periods without

energy supply (night time).

SAFETY

•     The unit must only be used by adults. This appliance is not intended for use by persons (including

children) with reduced physical, sensory or mental capabilities, or lack of experience and

knowledge, unless they have been given supervision or instruction concerning use of the appliance

by a person responsible for their safety. Children should be supervised to ensure that they do not

play with the appliance.

•     All installation work and adjustments to the unit must only be carried out by qualified personal.

Work performed by persons with insufficient technical knowledge may adversely affect the

performance of the unit or cause physical injury or damage to the equipment.

•     All servicing and repairs must only be carried out by a qualified customer service engineer. Only

genuine spare parts must be used.

•     Ensure that the connecting cable is not squeezed or bent when the unit is being installed or moved.
•     before cleaning or carrying out maintenance work, always switch the unit off and disconnect from

power supply.

       

WARNING : The unit's cooling system contains flammable refrigerant. Due to this fact, the

following information is particularly important.

•     Do not damage the cooling system! both the unit and the products stored inside can be severely

damaged if this cooling system starts to leak. Make sure, therefore, that no sharp or pointed objects

come into contact with the cooling system.

•     No electrical equipment may be operated inside the appliance.
•     The refrigerant in the appliance is an environment-friendly. before disposing of the appliance, the

refrigeration circuit should be opened outside so that the refrigerant can escape.

       Attention: The refrigerant is flammable!

TRANSPORT

•     Check whether the unit has been delivered undamaged. If you find that damage has occurred in

transit, immediately contact the delivery service or relevant sales outlet, submitting the delivery

note or proof of purchase. Do not operate a unit that has been damaged in transit! If you are unsure,

contact your sales outlet and ask them.

•     The unit must be transported in an upright position only (maximum inclination 45°).

ENVIRONMENTAL PROTECTION

•     Make your contribution to saving the environment. bear in mind that orderly and proper disposal is

required. Packaging materials and devices are always recyclable and should be taken for recycling.

•     before scrapping an old unit, remove the lids so that children cannot lock themselves inside while

playing.

•     When disposing of the unit, make sure that it does not get too hot, as combustible gas would cause

the insulating foam to froth up.

Содержание E003/043

Страница 1: ...ATION and MAINTENANCE INSTALACI N OPERACI N y MANTENIMIENTO INSTALLATION OP RATION et ENTRETIEN INSTALA O OPERA O e MANUTEN O TCW 2043 SDD SOLAR DIRECT DRIVEN REFRIGERATOR FREEZER PQS CODE E003 043 82...

Страница 2: ......

Страница 3: ...________________7 FREEZER COMPARTMENT ______________________________________8 USING ThE TCW2043SDD _______________________________________8 CONDENSATION WATER _______________________________________9...

Страница 4: ...tomer service engineer Only genuine spare parts must be used Ensure that the connecting cable is not squeezed or bent when the unit is being installed or moved before cleaning or carrying out maintena...

Страница 5: ...ice must be standing horizontally Use a water level to check and to correct if necessary The TCW2043SDD is equipped with bearing rolls Once it is in its final position the feet must be unscrewed to im...

Страница 6: ...ording to WhO specification Each generator must provide a voltage range of 12V to 20V DC with a minimum power of 320W at STC before connecting the TCW2043SDD and switching it on for the first time all...

Страница 7: ...ature setpoint STORAGE OF VACCINES IN ThE REFRIGERATOR IMPORTANT When using the TCW2043SDD refrigerator it is of utmost importance that the icelining waterpacks are in place Make sure that all waterpa...

Страница 8: ...s also applies if the equipment is switched off and on again Close the lid and allow the unit to run The temperature inside the cabinet is indicated on the display Observe the temperature from time to...

Страница 9: ...ked by a blue square Please refer to the logger user manual for further information CLEANING IMPORTANT Before cleaning the TCW2043SDD the vaccines preserved in it will have to be removed and stored te...

Страница 10: ...ILy Evacuate the condensation water as described before Take temperature readings and note down the temperature and the time of reading keep the temperature records systematically It is suggested that...

Страница 11: ...lem might occur if the solar generator doesn t deliver enough power cloudy day shadow on solar panels bad positioning of solar panels etc Check the actuals situation of the panels and if the problem p...

Страница 12: ...___________________16 COMPARTIMENTO DEL CONGELADOR_____________________________17 USO DEL TCW2043SDD_________________________________________17 AGUA DE CONDENSACI N ___________________________________...

Страница 13: ...piar o realizar trabajos de mantenimiento hay que desconectar la unidad siempre al igual que la alimentaci n el ctrica hay que tirar siempre al enchufe y nunca al cable Cuando en enchufe no est accesi...

Страница 14: ...y si es necesario corregir Esta unidad se suministra con ruedas para facilitar su transporte Situada en su emplazamiento final proceder a fijar la unidad por medio de los pies ajustables Para ello de...

Страница 15: ...erdo a la especificaci n de OMS Cada generador debe suministrar un rango de voltaje de 12V a 20V DC con una potencia m nima de 320W en STC Antes de conectar el TCW2043SDD y ponerlo en operaci n por pr...

Страница 16: ...el refrigerador TCW2043SDD es de suma importancia que el revestimiento de paquetes de agua se encuentre en su lugar Aseg rese de que todos los paquetes de agua est n colocados en su lugar antes de in...

Страница 17: ...3 minutos Esto se aplica a cada reinicio del TCW2043SDD Cerrar la tapa y esperar unos minutos a que la unidad se ponga en funcionamiento Comprobar la temperatura marcada de vez en cuando Inicialmente...

Страница 18: ...maci n LIMPIAR IMPORTANTE Antes de descongelar la unidad hay que quitar los paquetes de agua congelados situados en la misma y almacenarlos temporalmente en otro refrigerador Si no hay un segundo refr...

Страница 19: ...MIENTO DIARIO Evacuar el agua de condensaci n como se describe anteriormente Registrar los valores de temperatura del term metro anotando tambi n la hora de la lectura Guardar los registros sistem tic...

Страница 20: ...a unidad conectada Este problema puede ocurrir si el generador no suministra suficiente energ a d a nublado sombra en los paneles solares posici n err nea de los paneles solares etc Revise el estado d...

Страница 21: ...TEUR ________________25 COMPARTIMENT CONG LATEUR _________________________________26 NOTES IMPORTANTES POUR L UTILISATION DU TCW2043SDD ________26 EAU DE CONDENSATION__________________________________...

Страница 22: ...S A V comp tent N utiliser que des pi ces d tach es d origine Attention ne pas pincer ou tordre le c ble de raccordement pendant l installation ou le transport Avant le nettoyage ou la maintenance ne...

Страница 23: ...tement l appareil doit tre parfaitement de niveau Si n cessaire v rifier le niveau l aide d un niveau bulle Le TCW2043SDD est quip de 4 roulettes Quand l appareil est en place d visser les pieds l aid...

Страница 24: ...ation de l OMS Chaque g n rateur doit fournir une gamme de tension de 12V 20V DC avec une puissance minimale de 320W conditions de test standard Avant de brancher le TCW2043SDD et la premi re mise en...

Страница 25: ...Pour le fonctionnement correct du r frig rateur TCW2043SDD il est essentiel que les accumulateurs soient en place Assurez vous que tous les accumulateurs sont en place avant de d marrer le r frig rate...

Страница 26: ...ce le compresseur ne d marre qu apr s un temps de d c l ration de 3 minutes Ceci s applique apr s chaque remise en service Fermer le couvercle et laisser fonctionner l appareil Le thermom tre indique...

Страница 27: ...en option int gr e sur la connexion marqu e par un carr bleu Veuillez vous r f rer au manuel d utilisation de l enregistreur pour plus d informations NETTOYAGE IMPORTANT Avant le nettoyage les vaccin...

Страница 28: ...NANCE P RIODIQUE ChAQUE jOUR Evacuer l eau de condensation comme d crit ci dessus Relever plusieurs fois la temp rature indiqu e par les thermom tres et noter les valeurs et les heures Archiver syst m...

Страница 29: ...imentation lectrique Ce probl me peut appara tre si le g n rateur solaire ne produit pas assez d nergie temps nuageux mauvais positionnement des panneaux solaires etc Contr lez la situation des pannea...

Страница 30: ...DOR ________________34 COMPARTIMENTO DO CONGELADOR _____________________________35 UTILIZAR O TCW2043SDD_______________________________________35 GUA DE CONDENSA O ____________________________________...

Страница 31: ...genheiro de assist ncia qualificado Apenas devem ser utilizadas pe as sobressalentes originais Assegurar que o cabo de liga o n o est apertado ou dobrado quando a unidade est a ser instalada ou movida...

Страница 32: ...si o horizontal Utilizar um n vel de bolha de ar para verificar e se necess rio corrigir O TCW2043SDD est equipado com roletes Assim que estiver na sua posi o final os p s devem ser desenroscados para...

Страница 33: ...m2 dia de acordo com as especifica o da OMS Cada gerador deve assegurar um intervalo de varia o de voltagem de 12V a 20V DC com uma pot ncia m nima de 320W em STC Antes de conectar e ligar o TCW2043S...

Страница 34: ...FRIGERADOR IMPORTANTE Ao utilizar o TCW2043SDD como refrigerador extremamente importante que o revestimento de acumuladores de gua esteja corretamente colocado Certificar se de que todos os acumulador...

Страница 35: ...ap s uma demora de 3 minutos Isto tamb m se aplica no caso de o equipamento ser desligado e novamente ligado Fechar a tampa e permitir que a unidade trabalhe A temperatura do interior do recipiente e...

Страница 36: ...manual de utilizador do registador LIMPEZA IMPORTANTE Antes de proceder limpeza do TCW2043SDD as vacinas nele preservadas t m de ser removidas e temporariamente armazenadas num outro refrigerador Se n...

Страница 37: ...AMENTE Descarregar a gua de condensa o como descrito anteriormente Tirar as medi es da temperatura e apontar a temperatura bem como a hora de medi o Manter sistematicamente um registo da temperatura R...

Страница 38: ...Manter a unidade ligada Este problema poder ocorrer se o gerador solar n o fornecer alimenta o suficiente dia nebuloso sombra nos pain is solares mau posicionamento dos pain is solares etc Verificar a...

Страница 39: ...94 2863 01 Lid freezer cpl green 105 294 5104 06 Gasket silikon 14 5 18 L 2105 106 294 5104 05 Gasket silikon 14 5 18 L 1940 107 292 2516 00 Distance holder M8 17x60 108 296 4018 01 hinge protection t...

Страница 40: ...206 292 9225 10 Clamp Icepack holder 207 292 8580 09 Icepack TCW SDD 208 292 8513 01 Screw sealing cap 209 294 5406 32 Polystyren part 210 292 8513 00 Screw sealing cap Icepack 211 292 8508 01 Icepac...

Страница 41: ...92 Plate WAGO 307 292 2367 10 Clamp 308 296 9243 15 Fuse 15A 309 296 9804 74 Sensor Refrigerator 310 296 8954 85 Cable Set TCW2000SDD Refr 311 292 8517 19 Front mounting panel 312 296 3602 82 Control...

Страница 42: ...rtment 407 296 2026 03 Cable sleeve 408 296 2026 01 Nut for cable sleeve 409 292 8566 29 Small condenser 410 296 3577 00 bracket for small condenser 411 296 9960 06 Fan 412 292 8537 02 Fan bracket 413...

Страница 43: ...CTRICO SCH MA LECTRIQUE DIAGRAMA EL TRICO DC IN ST1 8 7 6 5 4 3 2 1 TCW DC FAN COMPR T C D F 3 2 1 3 2 1 ST1 8 7 6 5 4 3 2 1 TCW DC REFRIG FREEZER COMPR T C D F REFRIG FREEZER OUT IN 296 9769 77 296...

Страница 44: ...b Medical Systems S r l op der hei 17 L 9809 hosingen Luxembourg...