23
Step 3: Select the range.
Press Range and select the range. Auto range (Auto) can automatically select the appropriate range for
measurement based on the input signal. The auto range is adjusted upwards to 110% of the existing
range and down to 10% of the existing range.
Step 4: Set the integration time.
The setting of the integration time affects the measurement speed and measurement accuracy. The
longer the integration time, the higher the accuracy but the slower the measurement. PLC (power-line
cycle). 1PLC, 10PLC, 100PLC can inhibit the power supply noise. Selecting 100PLC can provide the best
noise rejection, but at a slower rate.
3.3.8 Capacitance Measurement
Step 1: Configure the test leads as follows:
Step 2: Press
on the front panel to enter the capacitance function interface.