background image

OPTIONS

Dosage Calculation

Vista

 basic Instructions for Use

16/25

950787 Rev 6/02

DOSAGE CALCULATION

This feature allows a rate to be calculated given the concentration and dose.

1.  Press OPTS until display shows DOSAGE CALC.
2.  Press ON.
3.  If concentration unit desired is other than the one displayed, press button

under the mcg? until units desired appear in the CONCENTRAT. (0*/ml) line.

4.  Enter the concentration of the solution.  If the concentration of the solution is

not known, go to #6 below. Press  C  to clear current data and re-enter new
data. 

5.  Press OK.
6.  To calculate the concentration of the solution, use 

 to move the * to the mg

field.  Enter the amount of drug in the solution.  Then press 

 to move the *

to the ml field.  Enter the amount of solution.

7.  Press OK.
8.  Enter the body weight if part of the dosing parameters, then press OK.  If

body weight is not part of the dosing parameters, press OK with 0 in the body
weight field.

9.  If time unit desired is other than the one displayed, press button under the

/h? until time unit desired appears in the first line. Enter the dose or use 

Æ

 to

move the * to the rate field and enter the rate.

10.  Press OK.
11.  Press

to begin the infusion.

To Review Data During Infusion

1.  Press INFO twice.

To Exit Dosage Calculation

1.  Stop the infusion.
2.  Press OPTS until display shows DOSAGE CALC.
3.  Press OFF.
4.  Press END.

Содержание Vista basic

Страница 1: ...Vista basic 950787 Rev 6 02 Instructions for Use Rx Only ...

Страница 2: ...ors can only be held responsible for any effects on device safety reliability and performance if Installation expansion work readjustments modifications or repairs are carried out by personnel authorized by the above The electrical wiring for AC power satisfies the requirements of VDE 0100 0107 and or the IEC publications or regional and or national variations The device is operated in line with t...

Страница 3: ...7 Total Infused Standby 18 Occlusion Pressure Drop Sensor Battery 19 Data Lock Contrast Selection 20 Adjust Clock Loudness 21 Alarms 22 AC 12V Battery CC Interface 23 Maintenance Storage Service 24 Specifications 25 The Vista basic is a volumetric infusion pump intended to provide accurate and continuous flow of parenteral and enteral fluids Qualified medical staff should decide how the device sho...

Страница 4: ...fully follow the instructions for loading removing and reloading the set as well as the recommended set change intervals Change pumping section on tubing every 24 hours Do not use tubing for more than 72 hours or per current facility protocol Always check pump data prior to starting infusion Do not operate this device in environments where there is an explosion hazard A possible explosion hazard e...

Страница 5: ...er it is possible there may be mutual influence Possible incompatibilities can be found in the instructions for use for the drug or for the other appliances Use only equipment accessories parts and disposables that are intended for use with the Vista basic Do not clean disinfect or sterilize any part of the device by autoclaving or with ethylene oxide gas Doing so may damage the device and void th...

Страница 6: ...displayed in plain language Green background lighting illuminates only when connected to AC power or in battery operation only when a key is pressed Keyboard Allows data entry Clear Key Resets a value to zero Alarm Light Lights up during an alarm Display will also show flashing AAA A Infusing Light Indicates pump is infusing ...

Страница 7: ...e 6 25 950787 Rev 6 02 Door Open Button Aluminum Case Easy upkeep drip proof and resistant to disinfectants Function Keys Allow choices of YES NO OK INFO OPTS etc ON OFF Power Key Alarm Silence Key Silences alarm tone for 2 minutes START STOP Key ...

Страница 8: ...according to the instructions provided on the set package Close the roller clamp 2 Check that the drip chamber is about 1 3 full and there is no air in the set 3 Open the pump door by pressing the Door Open Button on the top right corner of the pump 4 Push and hold open the clamp at the top of the fluid path 5 Insert tubing in clamp and release clamp 6 Thread the tubing down the fluid path making ...

Страница 9: ...red Press the green key below your answer YES Press YES to begin priming set Press STOP to end priming Press EXIT to return to the beginning of the last program with all data retained NO Returns to the beginning of the last program with all data retained CONTINUOUS MODE This section explains how to program a Continuous Mode infusion For directions on how to program other Modes and Options refer to...

Страница 10: ... Press to begin the infusion Check for drops in the drip chamber RATE and TIME 1 Enter rate in ml hr 2 Press TIME 3 Enter infusion time in hours and minutes 4 Press OK To leave screen without saving changes press EXIT 5 Press to begin the infusion Check for drops in the drip chamber To Change the Rate 1 Press to stop the infusion 2 Press The display will reset to 000 0 ml h 3 Enter a new rate 4 Pr...

Страница 11: ...hichever is less 1 Stop the infusion 2 Turn off the pump or reset the program INFORMATION KEY This key is only available while the pump is infusing 1 Press INFO 2 The Total Infused value and the Total Time of the infusion completed are displayed 3 Press INFO again and the Battery Left and Total Operation time of the pump is displayed 4 The pump will return to the main infusion screen after 10 seco...

Страница 12: ...r clamp on the Piggyback set 2 Hang the Piggyback solution container at least 8 inches above the primary solution container 3 Program the continuous infusion in the Continuous Mode see Continuous Mode 4 Press OPTS 5 Press ON to turn the Piggyback feature on 6 Press END PIGY will now be displayed in the lower left corner 7 Press PIGY 8 Enter the Piggyback infusion data as you would Continuous Mode ...

Страница 13: ...E 2 Press ON 3 Enter the total bag VOLUME to be delivered 4 Press OK 5 Enter the RAMP UP time in hours and minutes NOTE The pump defaults to 01h00m To change press and enter the time 6 Press OK 7 Enter the TAPER DOWN time in hours and minutes NOTE The pump defaults to 01h00m To change press and enter the time 8 Press OK 9 Enter the TOTAL Time for the infusion 10 Press OK Note The pump calculates t...

Страница 14: ...ue to program a new Ramp Taper program To Exit Ramp Taper Mode 1 When infusion is complete press to stop the KVO 2 Press EXIT 3 Press END to return to Continuous Mode Fast Taper The infusion can be interrupted to begin an early taper down process during the Ramp Taper Mode Infusion 1 Stop the infusion 2 Press FAST 3 Enter the FAST Taper Time in hours and minutes 4 Press OK 5 Press to start the inf...

Страница 15: ...s screen returns to Step 1 for review and changes 10 Press to start the infusion Check for drops in the drip chamber To Review or Change Program Mode data 1 Press OPTS until display shows PROGRAM MODE Note If Program Mode is infusing stop the infusion and press EXIT 2 Press ON 3 Review the data on each screen and press OK if data is correct or press to clear the current data and re enter new data ...

Страница 16: ...he infusion should begin NOTE This is NOT a clock time like 12 30 pm 5 Press OK 6 Enter how long in hours and minutes after the end of the infusion the pump should wait before alarming 7 Press OK 8 Press YES if the pump should run at KVO during the delay times Press NO if the pump should not infuse during the delay times 9 Review the data and press OK To change the data press EXIT and begin with 3...

Страница 17: ...e the concentration of the solution use to move the to the mg field Enter the amount of drug in the solution Then press to move the to the ml field Enter the amount of solution 7 Press OK 8 Enter the body weight if part of the dosing parameters then press OK If body weight is not part of the dosing parameters press OK with 0 in the body weight field 9 If time unit desired is other than the one dis...

Страница 18: ... right 2 Continue to hold down the first BOL while also pressing the second BOL A bolus will be given as long as both BOL keys are held down The pump will display the amount of bolus being administered and will beep once after each milliliter delivered INTERVAL BOLUS INTERMITTENT MODE The basal KVO rate and entire bag volume must be entered in CONTINUOUS MODE first 1 Press OPTS until display shows...

Страница 19: ...DBY This function allows the pump to be put on standby while retaining all set infusion data This function is only available in Continuous Mode 1 Press OPTS until display shows STANDBY 2 Press ON 3 Enter time in hours and minutes the pump is to stand by 4 Press C to clear the current data and re enter new data Press OK to accept new data 4 Press OK The pump will count down to zero and then alarm P...

Страница 20: ...ation without alarming DROP SENSOR Using a drop sensor eliminates the need for pre setting a VTBD in Continuous Mode and senses upstream occlusions and empty solution containers NOTE Turning the Drop Sensor feature off forces volumes to be entered and prevents sensing of upstream occlusions To Turn On the Drop Sensor 1 Press OPTS until display shows DROP SENSOR 2 Press ON 3 Press END to return to ...

Страница 21: ...tinuous Mode To Turn Off Data Lock 1 Press OPTS until display shows DATA LOCK 2 Press the decimal key 3 Press OFF 4 Press END to return to Continuous Mode CONTRAST SELECTION This feature allows the contrast of the screen to be changed for best possible viewing 1 Press OPTS until display shows CONTRAST SELECTION 2 Press to increase the contrast make the text darker Press to decrease the contrast ma...

Страница 22: ...ate using two digits for month day and year Example November 5 2001 is entered as 11 05 01 4 Press OK 5 Press TIME to enter the time of day Note The time is set using 24 hour Military Time Example 6 30pm is entered as 18 30 6 Press OK 7 Press END to return to Continuous Mode LOUDNESS This feature allows the volume of the alarm tone to be changed 1 Press OPTS until display shows LOUDNESS 2 Press to...

Страница 23: ...ion Close door Dwnstream Occlusion Downstream tubing kinked Filter blocked Closed clamp IV positional or infiltrated Catheter or vein too small Occlusion limit too low Correct kinks Correct blocked filter Open clamps Check IV site Increase occlusion limit Invalid data Entered data is out of range Press re enter data Keep Vein Open Rate Infusion Complete Programmed infusion is complete Stop infusio...

Страница 24: ...tery alarm or after 2 months of non use Charging time approx 16 hours Longer charging has no detrimental effect Full battery provides operation for 3 0 hours at highest delivery rate Battery capacity is approx 50 after 2 years The service life of the battery can be extended by completely discharging it from time to time and then fully recharging it on AC power CC Interface Interface descriptions a...

Страница 25: ...ng solutions Always turn off the pump and disconnect it from the electrical outlet prior to cleaning Storage Store the pump away from excessive heat cold or humidity Keep devices plugged into electrical power during storage Service If the pump fails to respond to the operating or troubleshooting procedures listed in this manual and the cause cannot be determined discontinue use and forward it to a...

Страница 26: ...5 measurement time 8hr temp 71 6 F Alarms Air in Line Low Battery Plug In Now Bag Empty Downstream Occlusion Door Open Pump Stopped KVO Unit Malfunction Air in Line Alarm Air bubble size 0 3mL or 1 5mL hr Can be set from 0 01 to 0 3 mL in service program Can be set from 0 5 to 3 5 mL hr in service program Memory User option to retain last Continuous and Piggyback Mode data upon power up Permanent ...

Отзывы: