Current range
0.00 – 5.00 mA; 0.00 – 1.00 mA
Stimulus duration
1.00 ms, 0.50 ms (optional), 0.30 ms,
0.10 ms, 0.05 ms (optional)
Frequency
2 Hz, 1 Hz, SENSe (3Hz)
Info
Battery, Date, Serial No.,
Version, Distributor, Manufacturer,
User Info Infrared
Setup
• Tone: Volume (0-8), mode
(click, beep), variable tone
• Dial Turns: 1,2
• Contrast: 0 – 8
• Auto Switch-off: 0 – 30
• Language: English, German, ...
• Options: Factory standard,
El. charge nC,
Auto adjust current
Add. stimulus duration
• Date: Hour, minute, ...
Menu structure:
Brief Instructions for Use: Stimuplex
®
HNS 12
Current threshold on/off
Configuring the basic settings for current range (example):
-> Switch on:
Display:
0.00,
Basic settings.
Confirm
selection
Up/down
Confirm
selection
Select Main
Menu
Up/down
Current range
Stimulus duration
Frequency
Current threshold
Setup
0.00 – 5.00 mA
0.00 – 1.00 mA
Info
-> Press right menu key, select current range (with up-down menu key),
-> Confirm current range with right menu key,
-> Select desired value with up/down key confirm with right menu key .
Right key: Select menu, confirm
Up/down key: Select desired value
Left key: One step back
Factory settings on delivery:
Maximum current 5 mA
Stimulus duration: SENSe
Stimulation
(0.10ms - 0.10ms - 0.15ms to 1.0ms)
Scale range:
SENSe (3Hz)
frequency:
Auto shut-off:
1 turn
After 20 Min.
If the stimulus duration-dependent current threshold
is out of tolerance, the target current is displayed
as contoured digits and the yellow LED blinks.
A warning signal is additionally sounded.
For procedure refer to reverse page.
Battery 9 V alkaline
mA key: Set current range
mA
ms key: Set stimulus duration
ms
Hz key: Set stimulation frequency
Hz
Stand-by key
MENU
MENU
MENU
MENU
MENU
SENSe
LLDorder 2641 - grunewald 150019
6043571_Stimuplex_HNS12_KurzAL_GB.indd 1
04.02.15 09:31