Chapter 6
OPERATION OF A PUMP WITH WIFI
6.2 Default configuration of the wireless battery:
SSID:
Space_xxxxxx (xxxxxx represents the serial
number of the pump with leading zero, e.g.
012345)
IP-Address (static): 192.168.101.41
Network type:
ad-hoc
Encryption:
none (open network)
6.3 Software applications of the wireless battery:
Space Online:
The wireless battery contains a web server which offers different
option to monitor and setup the wireless interface. Open a brow-
ser, i.e. Internet Explorer, and use the URL
http://ip_address_of_ wlan_pump, e.g. http://192.168.101.41
Attention:
The data, the values displayed on the web sites
and in particular the infusion pump alarms do not release from
observing the local alarm displays at the infusion pumps.
Therapeutic and diagnostic decisions must not be made solely on
the basis of the displays of the web application. SpaceOnline
offers different Micro-Sites upon which different data or
configurations are possible.
Data communication via BCC protocol:
Within SpaceOnline you can setup the BCC communication pro-
tocol. Please refer the BCC interface specification for further
details.
Attention:
Therapeutic or diagnostic conclusions must not
be based exclusively on values delivered by the BCC protocol and
their display on a Patient Data Management System or in the
web application. In particular the interpretation of alarms does
not release the operator from observing the local alarms at the