34
Alarms
9.2.1
Pre-alarms
In the event of a pre-alarm, an acoustic
signal sounds and a staff call is activated.
The display remains in pre-alarm until the
operating alarm goes off. Pre-alarms do not
cause delivery to be interrupted.
Display
notification
Meaning
“Volumes
nearly
infused”
Q
Preselected volume
has almost been
infused
Q
Remaining volume is
displayed
“Infusion
time nearly
reached”
Preselected time is almost
over
“Battery
nearly
empty”
The battery is almost
discharged
“KVO runs
for another
xx min:sec“
Volume/time have been
reached and the pump
continues with KVO rate.
A pre-alarm can be muted for 2 minutes by
pressing the
OK key
. The following symbol
is shown in the display:
9 Alarms
9.1
Device alarms
If a device alarm is triggered the infusion is
stopped immediately.
Q
Press the
On/Off key
to switch off
the device.
Q
Switch the device on again.
If there is another technical alarm:
Q
Disconnect the patient.
Q
Remove the disposable article.
Q
Switch off the device and send it to
the technical service.
9.2
Pre-alarms and operating
alarms
WARNING! Setting alarm thresholds
incorrectly may endanger the patient.
Q
Ensure that the alarm limits are set
so that the alarm can be triggered
in good time. This applies for maxi-
mum pressure in particular.
The operating alarms have a high priority.
Pre-alarms and reminder alarms have a
lower priority. If there are two pre-alarms
at the same time, the pre-alarm with the
shorter remaining time is displayed.
The time lag between the triggering of
the alarm and the activation of a staff
call is less than a second and is therefore
negligible.
The alarm pre-settings are retained in the
event of a power failure.