23
Set-up and start-up
7.3.5
Configuring service settings
Q
Select
Service...
and press
OK
to con-
firm.
Q
Enter the service code
and press
OK
to confirm.
The “Service Menu” screen is displayed.
Configuring the display language
Q
Select
Language
and press
OK
to con-
firm.
Q
Select the language
and press
OK
to
confirm.
Setting the date and time
Q
Select
Date
and press
OK
to confirm.
Q
Enter the day, month and year
and
press
OK
to confirm.
Q
Select
Time
and press
OK
to confirm.
Q
Enter the time
and press
OK
to
confirm.
Setting the bolus rate
Q
Select
Bolus rate
and press
OK
to
confirm.
Q
Set the bolus rate and press
OK
to
confirm.
Alarm
level
Pressure value
4
0.300 bar (225 mmHg)
5
0.400 bar (300 mmHg)
6
0.500 bar (375 mmHg)
7
0.700 bar (525 mmHg)
8
0.900 bar (675 mmHg)
9
1.000 bar (750 mmHg)
Note:
Occlusion must be resolved before
the infusion is re-started. Otherwise it will
affect the measurement accuracy.
The set pressure level is shown with a P
(for pressure) and a number. In addition,
a red area shows how quickly the set
pressure alarm limit will be reached. The
“manometer” display shows the current
pressure in the system. The lower the set
pressure alarm limit level is, the larger the
red area is, the quicker this limit is reached
and a pressure alarm triggered.