Aesculap
®
Detachable Rongeurs
Simply Clean
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany | Phone +49 7461 95-0 | Fax +49 7461 95-2600 | www.aesculap.com
Aesculap – a B. Braun company
C00511 0315/2.0/1
Open
n
Push
the golden button.
n
Pull
down the front handle.
n
Pull back the slider and
lift
it up.
Do not open the slider
more then 90°.
C
B
D
Close
n
Pull the slider right back and
place the slider on the main part.
n
Position the slider so that the
two arrows are congruent.
n
Push up the front handle.
n
Check that the rongeur
functions correctly.
Lubricate
Prior to function checks, lubricate
moving parts (especially the joints)
with maintenance oil suitable for
the respective sterilization process
(i.e. Aesculap Sterilit
®
JG598 or JG600
for steam sterilization)
Clean
n
Please follow the
instructions for use!
n
Keep the working ends
open for cleaning.
n
Place products in the tray
with their joints open.
Disassembly and Cleaning
Lubrication and Assembly
A
3
1
2
1
3
2