Aesculap

®

  

Detachable Rongeurs

Simply Clean

Aesculap AG  |  Am Aesculap-Platz  |  78532 Tuttlingen  |  Germany  |  Phone +49 7461 95-0  |  Fax +49 7461 95-2600  |  www.aesculap.com

Aesculap – a B. Braun company   

 

 

 

 

         

            C00511              0315/2.0/1

Open

Push 

the golden button.

 

Pull 

down the front handle.

Pull back the slider and 

lift

 it up.

 

Do not open the slider  

more then 90°. 

C

B

D

Close

Pull the slider right back and

  place the slider on the main part.  

Position the slider so that the

  two arrows are congruent.

Push up the front handle.

Check that the rongeur 

  functions correctly. 

Lubricate

Prior to function checks, lubricate  
moving parts (especially the joints) 
with maintenance oil suitable for  
the respective sterilization process  
(i.e. Aesculap Sterilit

®

 JG598 or JG600 

for steam sterilization) 

Clean 

Please follow the  
instructions for use! 

Keep the working ends  
open for cleaning.

Place products in the tray  
with their joints open. 

Disassembly and Cleaning

Lubrication and Assembly