Responsibility of the Manufacturer
The manufacturer, assembler, installer or
importer considers himself responsible for the
effects on safety, reliability and performance of
the equipment only if:
• assembly operations, extensions, re-adjust-
ments, modifications or repairs are carried
out by persons authorised by him,
• the electrical installation of the relevant
room complies with the appropriate require-
ments (e.g. VDE 0100, 0107 and/or the IEC-
publications resp. national requirements),
• the equipment is used in accordance with
the Instructions for Use and
• the Technical Safety Checks are carried out
regularly.
Warranty
B. Braun provides as from the date of delivery a
warranty of 2 years for every Infusomat® fmS.
This covers repair or replacement of parts dama-
ged as a result of design/manufacturing errors
or material defects. Modifications or repairs to
the unit undertaken by the owner or by third
parties invalidate the warranty.
The warranty does not cover the following:
Elimination of faults attributable to
incorrect/inexpert handling, or to normal wear
and tear and rechargeable batteries.
Technical Safety Check*
)
/ Service
The Technical Safety Check is recommended to
be carried out every 2 years and should be
documented.
Servicing work must be carried out exclusively
by personnel instructed by B. Braun.
Check regularly
Check for cleanliness, completeness and
damage. Use only according to Instructions for
Use. Check each time when switching on: self-
check, audible alarm, process- and alarm control
indication.
Cleaning
Clean using mild soap suds. Do not use spray
disinfectants at the mains power connection.
Recommended: disinfectant for wiping available
from B. Braun (e.g. Meliseptol®).
Before operation the device allow to vent for at
least 1 min. Do not spray into openings in the
device. Be sure to observe the instructions
provided concerning waste disposal and hygiene
for batteries and disposables.
18
Warranty / TSC
*)
/ Service / Cleaning
The CE mark confirms that this medical prod-
uct complies with the "Council Directive on
Medical Devices 93/42/EEC” dated 14
th
June
1993.
B. Braun Melsungen AG
Inspection on Delivery
Despite careful packaging, the risk of transport
damage cannot be entirely prevented. Upon
delivery, please check that nothing is missing.
Do not use a damaged device. Contact the
service department.
Items included
Infusomat® fmS, Power Cord, Drop Sensor, Pole
Clamp, Instructions for Use.