background image

Warning

Caution

• Never use dry heat sterilization, cold sterilization, etc.
•  Never autoclave the prophy angle handpiece together with others in 

a sterilization pouch.

An autoclave is recommended for sterilization.

•  Before sterilization, clean the surface of the prophy angle handpiece.
•  Lubricate the prophy angle handpiece, and then make a trial run.
•  Remove the prophy angle handpiece from the motor.
•  Put the prophy angle handpiece into a sterilization pouch when using 

a vacuum type autoclave. Do not combine it with other instruments in 
the same pouch.

•  Autoclave the prophy angle handpiece at 121 degrees C for 20 min-

utes or 134 degrees C for 15 minutes.

• Never use a wire brush or disinfectants for surface cleaning.
•  Never immerse the prophy angle handpiece in an ultrasonic cleaner 

or disinfectant solution.

•  Lack of lubrication after use or before autoclaving may cause prema-

ture failure of the prophy angle handpiece.

Sterilization

•  The prophy angle handpiece must be lubricated once a day or before 

autoclaving using the following procedures:

•   Use 

BA ULTIMATE SPRAY.

•  Remove the prophy angle handpiece from 

the motor. Take out brush or cup from the 
prophy angle handpiece.

•  Wipe off the saliva, dirt, etc. stuck to the sur 

face of the prophy angle handpiece with an 
alcohol-soaked soft tissue or cotton swab.

•  Keep the prophy angle handpiece in a 

horizontal position and inject BA ULTIMATE SPRAY for three seconds 
through the back-end of the prophy angle handpiece using the appro-
priate (E-Type) nozzle.

•  Install the prophy angle handpiece onto the motor after the spray 

injection.

•  Make a trial run of five seconds.
•  Wipe down the excess lubricant on the surface of the prophy angle 

handpiece with a soft tissue.

Lubrication

Caution

•  Never install/remove the prophy angle handpiece onto/from the motor 

while the motor is in operation.

•  Whenever abnormality in performance, sound level, heat level, func-

tion, etc. is sensed, stop using the prophy angle handpiece immedi-
ately and consult 

BA International Ltd. or an accredited BA certified 

Service Centre.

•  Never dismantle or alter the prophy angle handpiece.

Содержание Ultimate BA-102

Страница 1: ...o follow all the operating instructions including all cautions and warnings Keep this Operation Manual for future reference Install remove Brush Rubber Cup onto from prophy angle handpiece following t...

Страница 2: ...gle handpiece must be lubricated once a day or before autoclaving using the following procedures Use BA ULTIMATE SPRAY Remove the prophy angle handpiece from the motor Take out brush or cup from the p...

Страница 3: ...f the transmission gear engages with that of the head Engage the crown of the head with that of the shank and turn the union nut clockwise Removal Connection of Head from to E type shank Important Thi...

Страница 4: ...struc ciones de uso inclusive las precauciones y advertencias Conserve este manual de instrucciones como material de referencia en el futuro Instalaci n extracci n del cepillo de la tapa de goma en o...

Страница 5: ...una vez al d a o antes del autoclavado usando los siguientes m todos Utilice el ULTIMATE SPRAY de BA Extraiga la pieza de mano Prophy Angle del motor Extraiga el cepillo o la tapa de la pieza de mano...

Страница 6: ...n el del cabezal Encaje la corona del cabezal con la del huso y gire la tuerca de uni n en el sentido de las agujas del reloj Engage the crown of the head with that of the shank and turn the union nut...

Страница 7: ...note de toutes les instructions op rationnelles y compris des avertissements Conservez ce manuel d utilisation pour r f rence ult rieure Installer retirer la brosse cupule en caoutchouc sur de la pi...

Страница 8: ...Angle doit tre lubrifi e quotidiennement ou avant l autoclavage selon l une des proc dures suivantes Utiliser un spray BA ULTIMATE Retirer la pi ce main Prophy Angle du moteur Retirer la brosse ou la...

Страница 9: ...age de l engrenage de transmission s enclenche avec celui de la t te Enclencher la couronne de la t te avec celle de la tige et tourner le filetage dans le sens horaire Retrait connexion de la t te de...

Страница 10: ...ienungsanweisungen mit den Vorsichts und Warnhinweisen Bewahren Sie diese Bedienungsanleitung zu Ihrer Information auf Aufsetzen Abziehen der B rste des Gummikelchs auf das von dem Prophy Angle Handst...

Страница 11: ...terilisation Das Prophy Angle Handst ck muss einmal t glich oder vor dem Au toklavieren mit folgenden Verfahren gegl ttet werden Mit dem BA ULTIMATE SPRAY Entfernen Sie das Prophy Angle Handst ck vom...

Страница 12: ...e Rastnase im Kopf ein rastet Verbinden Sie den Kopfaufsatz mit dem Schaft und drehen Sie dann die Anschlussmutter im Uhrzeigersinn Abziehen Aufsetzen des Kopfes vom auf den Schaft Typ E Wichtig Diese...

Страница 13: ...assi Seguire attentamente le istruzioni d uso incluse le precauzioni e le avvertenze Conservare questo manuale per consultazioni successive Installare rimuovere lo spazzolino coppetta di gomma sul dal...

Страница 14: ...ficare il contrangolo per profilassi una volta al giorno o prima dell autoclavazione seguendo le procedure indicate di seguito Usare BA ULTIMATE SPRAY Rimuovere il contrangolo per profilassi dal motor...

Страница 15: ...della trasmissione si incastri con quello della testina Incastrare la corona della testina con quella dell anello e ruotare il dado di accoppiamento in senso orario Rimozione collegamento della testin...

Отзывы: