background image

Instructions for use

BA Optima

Contra-angle handpiece

BA60E (BA611075)

Содержание Optima BA60E

Страница 1: ...Instructions for use BA Optima Contra angle handpiece BA60E BA611075 ...

Страница 2: ...ration attachment detachment 10 Contra angle handpiece 5 Starting operation Changing the rotary instrument 11 14 Rotary instruments contra angle handpieces test run 6 Hygiene and maintenance 15 23 General pre disinfection manual cleaning manual disinfection machine cleaning lubrication sterilization and storage ...

Страница 3: ...pare parts 24 8 Technical data 25 26 9 Recycling and disposal 27 Explanation of warranty terms 28 Authorized B A International Ltd service partners 29 Symbols 30 31 in the instructions for use contra angle handpiece on the packaging ...

Страница 4: ...For your safety and the safety of your patients Please read the instructions for use before using the device for the first time They explain how to use the product and guarantee trouble free economical and safe operation Intended use The contra angle handpiece is used to clean and polish the surface of the teeth and fillings Misuse may damage the contra angle handpiece and thus cause risks and haz...

Страница 5: ...EEC Responsibility of the manufacturer The manufacturer can only accept responsibility for the safety reliability and performance of the contra angle handpiece if it is used in compliance with the following directions The contra angle handpiece must be used in accordance with these instructions for use The contra angle handpiece has no components that can be repaired by the user Modifications or r...

Страница 6: ...fection prevention self protection and patient safety Improper use e g due to lack of hygiene or appropriate maintenance non compliance with our instructions or the use of accessories and spare parts which have not been approved by B A International Ltd will invalidate all claims made under warranty or otherwise Servicing In case of malfunction please contact an authorised B A International Ltd se...

Страница 7: ... filtered oil free and cooled air supplied by dental compressors for drive air Always ensure that the correct operating conditions are provided Inspect the contra angle handpiece for damage and loose parts before every use Do not operate the contra angle handpiece if it is damaged Only fit the contra angle handpiece when the motor is at a complete standstill Perform a test run prior to each treatm...

Страница 8: ...fore using for the first time The contra angle handpiece is supplied in a clean condition sealed in PE foil Lubricate the contra angle handpiece Sterilize the contra angle handpiece Oil service page 20 21 Sterilization page 22 23 ...

Страница 9: ...9 3 Product description BA60E Thread Drive Spanner Sheath ...

Страница 10: ...ent Contra angle handpiece Do not attach or detach the contra angle handpiece while in operation Fit the contra angle handpiece on the motor Check that it is securely attached to the motor Remove the contra angle handpiece by pulling it axially ...

Страница 11: ...rded after use Only use fault free rotary instruments Observe the manufacturer s instructions Only insert the rotary instrument when the contra angle handpiece is static Never touch a rotary instrument that is still rotating Only use the contra angle handpiece in forward operation The rotary instrument can come loose in reverse operation risk of injury 2 ...

Страница 12: ...12 The contra angle handpiece head is equipped with a Triple Seal For caps brushes please contact your dealer ...

Страница 13: ...y cap brush Locking the contra angle handpiece drive Insert the spanner in the drive dog Lock the contra angle handpiece drive a Screw the cap brush in as far as it will go b Check that it is securely attached c Unscrew the cap brush b a c ...

Страница 14: ...at eye level Insert the rotary instrument Start the contra angle handpiece In the event of malfunctions e g vibrations unusual noises overheating stop the contra angle handpiece immediately and contact an authorized B A International Ltd service partner ...

Страница 15: ... the rotary instrument Clean and disinfect the contra angle handpiece immediately after each treatment to rinse out any liquid e g blood saliva etc that may have entered and to prevent the inside parts from seizing After manual cleaning disinfection and lubrication you must carry out a thermal disinfection unwrapped or sterilization wrapped in a class B or S steam sterilizer as per EN 13060 Steril...

Страница 16: ...16 Pre disinfection If heavily soiled clean first with disinfectant cloths Only use disinfectants which have no protein fixing effect ...

Страница 17: ...e and outside Rinse and brush off with demineralized water 38 C Remove any liquid residues absorbent cloth blow dry with compressed air Do not place the contra angle handpiece in disinfectant solution or in an ultrasonic bath ...

Страница 18: ...ternational Ltd recommends wiping disinfection Use only disinfectants which do not contain chlorine and which are certified by officially recognized institutes Note the manufacturer s specifications for the use of the disinfectant ...

Страница 19: ...isinfector Follow the manufacturer s recommendations for devices and cleaning rinsing agents Ensure that the contra angle handpiece is completely dry both inside and out following thermo washer disinfection Remove any liquid residues with compressed air Lubricate the dry contra angle handpiece immediately after thermo washer disinfection ...

Страница 20: ... Ultimate Spray Follow the instructions on the oil spray can and on the packaging Recommended maintenance cycles After every internal cleaning without fail Before every sterilization or After 30 minutes of use or At least 1x a day ...

Страница 21: ...the contra angle handpiece head downwards Allow the contra angle handpiece to run for 30 seconds in order to remove any excess oil Repeat the entire hygiene and maintenance process if any dirt escapes Wipe the contra angle handpiece with a cellulose cloth or soft cloth ...

Страница 22: ...he contra angle handpiece and the accessories in sterile goods packaging according to EN 868 5 Make sure that you only remove dry sterile goods from the sterilizer Store sterile goods dust free and dry ISO 14457 stipulates a durability of at least 250 sterilization cycles In the case of the B A International Ltd contra angle handpiece we recommend you have a regular service carried out after 1 000...

Страница 23: ... with sterilizers in accordance with EN 13060 Sterilization holding time minimum of 3 minutes at 135 C or Steam sterilization class S with sterilizers in accordance with EN 13060 The sterilizer manufacturer must expressly approve the sterilization of contra angle handpieces Sterilization holding time minimum of 3 minutes at 135 C ...

Страница 24: ...nternational Ltd accessories and spare parts or accessories approved by B A International Ltd Reference source B A International Ltd partners BAOIL B A Ultimate Spray 2038200 Oil spray cap with nozzle for straight and contra angle handpieces 1957300 Spanner ...

Страница 25: ...8 Technical data Contra angle handpiece BA60E Gear ratio 4 1 Outer diameter of the sheath mm 20 Motor connection as per standard ISO 3964 Maximum drive speed min 1 5 000 10 000 min 1 revolutions per minute ...

Страница 26: ...piece on the patient side maximum 50 C Temperature of the working instrument rotary instrument maximum 41 C Physical characteristics Temperature for storage and transport 40 C to 70 C Humidity for storage and transport 8 to 80 relative non condensing Temperature in operation 10 C to 35 C Humidity in operation 15 to 80 relative non condensing ...

Страница 27: ...ments Follow your country specific laws guidelines standards and directives concerning disposal Ensure that the parts are not contaminated on disposal Disposal of the packaging material All packaging materials have been selected according to environmentally compatible and disposal aspects and can be recycled Please send old packaging materials to the relevant collection and reprocessing system Thi...

Страница 28: ... for material or manufacturing defects within a warranty period of 12 months from the date of purchase We accept no responsibility for damage caused by incorrect handling or by repairs carried out by third parties not authorized to do so by B A International Ltd Claims under warranty accompanied by proof of purchase must be sent to the vendor or to an authorized B A International Ltd service partn...

Страница 29: ... A International Ltd service partners You can find your nearest B A International Ltd service partner at www bainternational com under the service section Alternatively please contact us at service bainternational com ...

Страница 30: ...tructions for use WARNING in cases where people could be injured ATTENTION in cases where something could be damaged General explanations without risk to persons or objects Do not dispose of with domestic waste Not for re use ...

Страница 31: ... packaging Catalogue number Serial number Thermo washer disinfectable Sterilizable up to the stated temperature Date of manufacture CE 0476 from the manufacturer Data matrix code for product identification e g in the hygiene and care process SN REF ...

Страница 32: ...Form Nr 50922 AEN Rev 000 19 07 2016 Subject to change Manufacturer B A International Ltd Unit 9 Kingthorpe Business Centre Studland Road Kingsthorpe Northampton NN2 6NE U K Rev2016 07 ...