• Hold the spray can upright.
• Be sure to hold the handpiece firmly to prevent slipping out of
hand by the spray pressure when lubricating.
CAUTION
Lubrication Oil
1) Remove the motor from the hose.
2) Apply 1-2 drops of lubrication oil into the Drive Air Port shown in
the Fig. 8.
3) Connect the motor to the hose. Operate the motor for about 10 seconds.
9-5 Sterilization
Sterilize the product by autoclave sterilization. After each patient
sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Autoclave for more than 15min. at 132°C, or 3min. at 134°C.
3) The handpiece should remain in the autoclave pouch until required
for use.
9-3 Disinfection
Manual Disinfection (External Surface)
Wipe the surface with cloth soaked in disinfectant solution such as
alcohol.
Automatic Disinfection (External and Internal Surfaces)
Use a thermo-disinfector in accordance with ISO 15883 (EN ISO 15883).
Refer to the thermo-disinfector manual.
9-4 Lubrication
BA ULTIMATE SPRAY
Apply BA ULTIMATE SPRAY every time after each use and/or before
autoclaving.
1) Remove the motor from the hose.
2) Mount the arrow-head spray nozzle tip into the spray can port.
(Fig. 7)
3) Insert the Spray Nozzle into the Drive Air Port of the motor. Hold the
motor and spray for approximately 2-3 seconds. Apply lubricant
until it expels from the exhaust air part for at least 2 seconds. (Fig.8)