CAUTION
•
Hold the spray can upright.
•
Be sure to hold the handpiece firmly to prevent slipping
out of hand by the spray pressure when lubricating.
•
Supply lubricant until it comes out of the handpiece
head.
•
If you wish to expel excessive oil from inside of the
handpiece, rotate the handpiece for approx.15 seconds
without a Rubber Cup / Brush.
WARNING
• If blood infiltrates inside a handpiece, an
automatic handpiece cleaning and lubrication
system may not totally clean the internal
handpiece components, which may lead to
internal blood coagulation. Coagulated blood
inside the handpiece may cause handpiece
failure and overheating, causing burn
injuries. To avoid the risk, B.A. International
recommends to immediately lubricate those
handpieces with BA ULTIMATE SPRAY.