background image

TEMPERATURE DATALOGGER

o

F

-22

~188

o

-30~70 C

SPECIFICATIONS

TEMP. RANGE

HUMIDITY  RANGE

        N/A

8813

8815

1K/2K/4K/8K

SIZE:12 x 25.5 mm

o

o

+

Temp.  0.6 C(-20~50 C)      

o

o

o

+

            1.2 C(-30~-20 C,51~70 C)

-

-

-

CR2032 x1

POWER BAT.

FRONT PANEL

ACCURACY

BAT LOW DISPLAY

"Lo" DISPLAY

SAMPLE DATA

SAMPLE POINTS

LCD DISPLAY

HOUSING

WATERPROOF

WATERPROOF 

DIMENSIONS(mm)

85(L)x61(W)x20.2(T)(USB type Interface)

124(L)x92(W)(Complete set)

LED SIGNAL

Red (HI,LO Alarm) /Yellow(Record)

RESOLUTION

CABLE LENGTH   

 

150 cm with DB9 connector

o

o

0.1 C (0.1 F) 

85(L)x61(W)x20.2(T)

40

8814

NO DISPLAY

                                      Up to 8000

Up to 4000

Up to 4000

o

C

o

C

o

C

Содержание 88128

Страница 1: ...OPERATION MANUAL TEMPERATURE HUMIDITY DATA LOGGER Model 88128 8828 8829 8813 8814 8815 8833 8834 8835 Interface...

Страница 2: ...setting A Sample point setup B Sample rate C Start mode set up D Logger s clock setting E Alarm setting F Unit setting Logger ID setting G Sleep Mode setting Data Table Data Table Edit Go to Find Sta...

Страница 3: ...ware is used to readout the stored data and see as a graph showing the history with real time clock the tabular data can be viewed or exported to a spreadsheet for various analysis One interface can b...

Страница 4: ...cable software Model 8824P White The USB Interface with cable software Model 88128S 8828S 8829S 8813S 8814S 8815S 8833S 8834S 8835S The complete set including data logger with battery interface and so...

Страница 5: ...rear upper for hanging on the wall to prevent losing or falling down One Interface can be used with several loggers to save cost Download data to other program for further analysis All data is saved i...

Страница 6: ...rm designed as not audible but only for reminding or warning the user to do necessary action REC per sample rate setting For 88128 if the interval is over or 8828 8829 if the interval is over 5 second...

Страница 7: ...e in the logger unit RH Display Relative Humidity Though you might have seen RH when replacing with a new battery which is not available for the model 8828 but model 8829 Temperature Humidity datalogg...

Страница 8: ...e steps to remove and replace battery 1 Unscrew the datalogger from the rear side 2 Do not remove o ring and make sure the o ring is in its place groove 3 Remove the expired battery 4 Insert a new bat...

Страница 9: ...o the computer for starting software set up a Install Software To install the program b Manual PDF c Open the PDF format manual Web Site Visit our website in your browser d Brow CD ROM Browse the CD R...

Страница 10: ...e left lower part cursor status changed while pressing arrow left or down arrow key for last record or pressing right or up arrow key for next record MAIN SCREEN Diagram A Main Screen LOGO Display Con...

Страница 11: ...ffset Knob after pressed Offset Knob in the following ways Press the up or down arrow key to increase or decrease one record number or corresponding to date in the graph FAQ Why can t I see the moveme...

Страница 12: ...ill see the number on the knobs of Offset and Range be changed The control range depended on each logger Logger Offset Range 8813 8814 0000 1000 2000 3000 50 250 500 1K 2K 4K 8815 8833 34 0000 2500 50...

Страница 13: ...grid can be drawn on the graph and a dialogue box display at the left bottom corner shows the appearance of this grid CONTROL LOCATION Cursor 1 and Cursor 2 11 Diagram B Retrieve File Look in Directo...

Страница 14: ...33 8834 you will not see RH on the right side of the screen but only model 8829 8835 See Dia B 2 However if you plug external probe on 8833 and 8834 you will see two different temperature curve Meanti...

Страница 15: ...cting icon select folder and name to save the data to be saved The windows Save File dialog box allows you to specify the file format to save the data the file name to be called where the file to be s...

Страница 16: ...t OK to accept setting changes Cancel to abort and to exit Setting menu Note Always remember to select 9600 baud rate for all models Data bits always is 8 Parity is None Stop bits is 1 Diagram D COM P...

Страница 17: ...TTING Diagram E Logger Setting Sample point Setup SAMPLE POINTS SETTING Select sample point 1 000 2 000 4 000 8 000 or 12 000 16 000 Note Only 88128 8828 8829 8835 are available for selecting sample p...

Страница 18: ...op logging Note The download operation before completing all sample points won t work for the logger sample rate to be set as 1 second 2 For 8813 14 15 and 8833 34 35 press start stop key for over 3 s...

Страница 19: ...t the mode as Key Start and press button again When set the mode as On line and press OK button the logger will start recording immediately and export data to computer synchronously So when choose thi...

Страница 20: ...used to have alarm setting before then the default will be the same as your last setting You could choose to change it or not User simply select desired High value or Low value by scrolling up for in...

Страница 21: ...is available in C and F See Diagram G UNIT SETTING Logger s ID Setting enables user to give unique ID names to each data logger up to 8 digits simplifying data tracking LOGGER ID SETTING 19 Please be...

Страница 22: ...printer or a fax internet application depending on your available facilities The print out points range is selectable See Diagram H Depress icon to view detailed data table which includes 1000 sample...

Страница 23: ...001 6 29 2001 6 29 2001 8 28 20 8 28 24 8 28 23 8 28 22 8 28 21 8 28 53 8 28 52 8 28 54 8 28 55 8 28 56 8 28 57 8 28 58 9 3 9 3 9 3 9 3 9 3 9 3 9 3 9 3 9 3 9 3 9 3 9 3 9 3 29 7 29 7 29 7 29 7 29 7 29...

Страница 24: ...in the Find What block Select from the following 1 Case Sensitive 2 While Cell 3 Search Vertically 4 Select Cell Only 5 Wrap Click any of above and press Find Next button to find press again to find t...

Страница 25: ...emperature RH external temperature 2 Condition AND OR 3 Value certain reading 4 Value certain reading 5 Error If selecting this all the Error code appeared in data table will be marked and counted 6 A...

Страница 26: ...24 Usage Records The software could help to count how many records is within the range you set Counter Times Help to count how many runs the data out of the select range and then come back again...

Страница 27: ...to see Diagram I 2 is another example of setting the 2 As shown in the middle top of Diagram I 1 by changing the cursor position on the graph you will find an example for the statistic screen of temp...

Страница 28: ...ME 11 23 2001 17 12 57 0001 0401 0801 1201 1601 2001 2401 2801 3201 3601 4000 4400 4800 5200 5600 6000 6400 6800 7200 7600 8000 08 03 2001 14 46 05 TEMPERATURE CHART Record No Time 08 03 2001 19 12 43...

Страница 29: ...e between two selected cursors Std Dev This is a very useful feature to see more reliable deviation while recording Calculate each deviation between each value and Mean value then get an average figur...

Страница 30: ...x internet program depending on your available facilities See diagram J Selecting Print button on the menu bar Windows Print dialog box allows you to change the pre set format then print the contents...

Страница 31: ...or of curve to refer to the file you retrieve by clicking the color block You can print the screen datas by clicking the printer icon at the right top corner 29 FAQ If I want to retrieve another file...

Страница 32: ...6198 6398 6598 6798 6998 71 B Date 07 16 2001 Time 15 26 10 T 17 0 RH 47 8 F Date 07 17 2001 Time 14 09 15 T 26 1 RH 67 4 Retrieve B file completed F TEMP F RH B TEMP B RH Display Control Example Grou...

Страница 33: ...the offset and range setting desired by moving the indicator on the yellow button Please see Diagram L refer to the introduction DISPLAY CONTROL 31 CURSOR STATUS The Cursor Status indicates exact figu...

Страница 34: ...s you follow up above instruction BLANK LOGO This area is designed to insert customers own logo The logo will be set as default image and can t be changed by the users Please refer to Diagram M 1 If p...

Страница 35: ...eans if you don t want to download complete sample readings you can always stop it at any time you want Exit To exit Download Commands DISPLAY CONTROL Offset To set up from which sample Reading you wo...

Страница 36: ...and dragging the red indicator to 4000 Range unchanged at 8K you will see the Diagram Q below The screen shows the number 4000 as starting point of the total 8 000 records See diagram Q 1 Actually th...

Страница 37: ...are different range for selection based on the logger Display Control For X Axis records For example Set Range by clicking dragging the red indicatoras Diagram R to 1K Diagram R 1 is shown the data r...

Страница 38: ...rd from 0001 to 4001 36 Diagram S 1 Diagram R 1 Logger Offset Range 8813 8814 0000 1000 2000 3000 50 250 500 1K 2K 4K 8815 8833 34 0000 2500 5000 7500 50 250 500 1K 2K 4K 8K 88128 8828 0000 5000 10000...

Страница 39: ...See Diagram T 1 There are Temp Temp2 and all three functions for selection Temp air temp Temp2 external temp All Temp Temp2 Above is the display of All Y scale will be decided by two temp For example...

Страница 40: ...QUARE ACCURACY SAMPLE DATA Up to 8000 up to 16000 SAMPLE POINTS HOUSING WATERPROOF 8818 Phased out 88128 DIMENSIONS mm 124 L x92 W x37 T Interface 108 L x66 W x22 T 124 L x92 W Complete set LED SIGNAL...

Страница 41: ...1 2 C 40 20 C 51 85 C Humidity 3 RH ER3 or CR2 lithium battery x1 POWER BAT FRONT PANEL ACCURACY BAT LOW DISPLAY Lo DISPLAY SAMPLE DATA Up to 16000 SAMPLE POINTS LCD DISPLAY HOUSING WATERPROOF WATER R...

Страница 42: ...OWER BAT FRONT PANEL ACCURACY BAT LOW DISPLAY Lo DISPLAY SAMPLE DATA SAMPLE POINTS LCD DISPLAY HOUSING WATERPROOF WATERPROOF DIMENSIONS mm 85 L x61 W x20 2 T USB type Interface 124 L x92 W Complete se...

Страница 43: ...hium battery x1 POWER BAT FRONT PANEL ACCURACY BAT LOW DISPLAY Lo DISPLAY SAMPLE DATA Up to 8000 SAMPLE POINTS LCD DISPLAY HOUSING WATERPROOF DIMENSIONS mm 85 L x61 W x20 2 T Interface 124 L x92 W x50...

Страница 44: ...er maintenance or damage resulting from leaking batteries Proof of purchase is required for warranty repairs RETURN AUTHORIZATION 42 FAQ If I can t down load and the display won t active what should I...

Страница 45: ...ter Anemometer Sound Level Meter Air Flow meter Infrared Thermometer K type Thermometer K J T type Thermometer K J T R S E type Thermometer pH Meter Conductivity Meter T D S Meter D O Meter Saccharime...

Отзывы: