![Axonics Model 1601 Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/axonics/model-1601/model-1601_manual_3040486015.webp)
15
• If there is no flashing light when you press the button, try pressing the button again. Confirm you are pressing the button as
shown in Figure 2. If there is no light after multiple attempts at pressing the button, do not use this TS. Activate a new TS. If the
issue persists with a new TS, contact Axonics.
Figure 2:
Activate the TS by pressing the button on the back of the stimulator
Connecting the Trial Stimulator to the Tined Lead or PNE Lead
The following section describes the process for connecting theTS to the cables used to connect to the tined lead or the PNE lead.This
should be performed when an Axonics lead has already been implanted and the appropriate cables are available. For a trial using
the PNE lead, the BasicTrial Cable should be connected to theTS. For a trial using the tined lead, a Percutaneous Extension (PE) cable
should be connected to the TS.
1. Align the raised base on theTS connector with the raised bar on the BasicTrial Cable or the grey colored bar on the PE (
Figure 3
).
2. Press the connectors together to connect the Basic Trial Cable or PE to the extension of the TS (
Figure 3
).
3. Use the CP to check the TS impedances to verify the TS and cable is connected correctly.
Notes:
• There will not be any gap between the connectors when fully inserted.
• Use the CP to check the electrode impedances to confirm that the cables are connected.
110-0049 rev G_output.indd 15
11/20/18 7:08 AM
Содержание Model 1601
Страница 2: ...2 ...
Страница 4: ...4 ...
Страница 20: ...20 ...
Страница 40: ...40 ...
Страница 58: ...58 ...
Страница 60: ...60 ...
Страница 78: ...78 ...
Страница 80: ...80 ...
Страница 98: ...98 ...
Страница 100: ...100 ...
Страница 118: ...118 ...
Страница 119: ...119 ...