Axium Clinical Programmer Скачать руководство пользователя страница 7

9

10

Programmer Status Bar

Located at the bottom of the Clinical Programmer screen, the Programmer Status Bar displays:

•   

Programmer-Stimulator Connection Status: 

Displays the status of the communication between 

the Clinical Programmer and the NS device: “Connecting” is displayed when establishing 

a connection. “Connected” is displayed when there is communication between the Clinical 

Programmer and the NS device. “Disconnected” is displayed when there is no communication 

between the Clinical Programmer and the NS device.

•   

Programmer Battery Level:

 Displays the Clinical Programmer battery charge level. It is 

recommended that the Programmer be plugged in and charging when not in use. Plug in and 

charge the Programmer before reaching 30% remaining life.

•   

Programmer Clock:

 Displays the time. See User Manual section on Change the Date and Time. 

Workspace Navigation

Once the NS is connected, tabs are displayed for 

the systems’ four main workspaces (“Profile”, “Stim”, 

“Group”, and “Utility”). The Workspaces are used 

to view and program the NS therapy settings and 

to obtain diagnostic information. A record of the 

programmed settings and diagnostic information is generated after every session. A fifth Workspace 

labeled “FCE” will only appear when FCE mode is ON.

Workspace screens and sub-screens are navigated by selecting the labeled tabs. 

Located at the bottom of each of the Workspaces are 

the “ALL”, “Program” and “Exit” buttons.

•   

Exit button: 

is used to close the current window, end the patient therapy session, and return to 

the Main Menu.

•  

 Program  button:

 programs all changes made within the current Workspace.

•  

 ALL  button:

 turns all stimulation off. 

NOTE: 

Returning to the Main Menu or turning off the Programmer will not change any of the 

programmed NS settings.

NOTE: 

When programming is complete, select the “Exit” button to conserve power.

Temporary and Permanent Programming

Whenever a change is made to a parameter value or other data field while the NS is within telemetry 

range, this value immediately becomes 

temporarily active

. The corresponding value or data selection 

appears in a red bold underlined font.

NOTE:

 The “temporarily active” state does not apply to the Group workspace. 

Temporary programmed values or text data can be 

permanently programmed

 by pressing the program 

button. The font color changes from red to black.

NOTE:

 When leaving a Workspace while values are 

temporarily active

 you will be prompted to 

either program these values permanently or cancel the pending changes. 

NOTE:

 Parameters can be temporarily active on multiple tabs of the same Workspace. 

More on Editing Text Fields

NOTE:

 Selecting a text field will pop up a keyboard at the bottom of the screen. To close the 

keyboard after modifying the entry, press the keyboard key centered in the blue bar at the bottom 

of the screen.

While text fields are 

being edited

, they appear in a black bold 

font (no underline). At the same time to the right of the text field a 

red dot indicates that editing is in progress. 

Once editing for a field is complete, tap the red dot to make the 

change 

temporarily active

. The red dot disappears and the font 

changes from black bold to red bold underlined. 

Only upon pressing the programming button does the change 

become 

permanently programmed

 and the font color changes 

from red to black.

Using the Workspaces 

Profile Workspace

Press the “Profile” tab to access the Profile Workspace. The Profile Workspace is divided into five tabs 

(“Patient”, “Clinic”, “NS”, “Leads” and “System”) which are used to:

•  Enter patient information
•  Enter clinician contact information
•  Enter NS device information
•  Enter lead identification information
•  Change basic system parameters

Содержание Clinical Programmer

Страница 1: ...thorized duplication or use is prohibited AXIUM SPINALMODULATION and the Man in Circle Design are trademarks and or registered trademarks of Spinal Modulation Inc Clinical Programmer Model MN20700 03 User Manual Authorization to affix the CE mark granted in 2011 ...

Страница 2: ... Workspace Navigation 9 Temporary and Permanent Programming 10 More on Editing Text Fields 10 USING THE WORKSPACES 10 Profile Workspace 10 Patient Information Tab Profile Patient 11 Clinic Information Tab Profile Clinic 12 Stimulator Information Tab Profile NS 12 Leads Information Tab Profile Leads 13 System Information Tab Profile System 13 Stim Workspace 14 Stim Tabs Stim Target Name 14 Impedanc...

Страница 3: ...ds and stimulator hardware used differ the programmer hardware and instructions for programming the TNS and INS devices are the same NOTE In this manual the general abbreviation NS is used for information which applies to both TNS and INS In all other cases the specific abbreviations TNS or INS are used For specific description of the TNS and INS system components and implant procedures refer to t...

Страница 4: ... concern regarding the proper function of the Spinal Modulation NS System please contact your Spinal Modulation representative RF OPERATING FREQUENCIES Nearby equipment emitting strong magnetic fields can interfere with RF communication even if the other equipment complies with CISPR emission requirements The operating characteristics are as follows MICS band 402 405 MHz The effective radiated pow...

Страница 5: ...cycles with normal use Connect the Clinical Programmer to the Charger and attach to an outlet regularly to keep it charged Programmer Power Up Turn the Clinical Programmer ON by pressing the button The Main Menu will be displayed NOTE If the Clinical Programmer screen does not turn on follow the instructions for charging the battery and try again Main Menu The Main Menu displays three primary func...

Страница 6: ... telemetry connection is cancelled 6 Move the Clinical Programmer magnet over the NS device in a circular motion to connect The indicator status bar on the bottom left of the screen will display Connected if the connection attempt is successful If the Programmer could not communicate with the NS device an error message will appear and Disconnected will be displayed in the status bar NOTE If after ...

Страница 7: ...e programmed NS settings NOTE When programming is complete select the Exit button to conserve power Temporary and Permanent Programming Whenever a change is made to a parameter value or other data field while the NS is within telemetry range this value immediately becomes temporarily active The corresponding value or data selection appears in a red bold underlined font NOTE The temporarily active ...

Страница 8: ...ng the áü button near the space bar allows the use of accented characters Clinic Information Tab Profile Clinic Enter or modify the physician and clinic information in the text fields provided Physician Name Clinic Name Clinic After Hours Contact Phone Number Clinic Phone Number Clinic Email Clinic Address Stimulator Information Tab Profile NS The NS tab provides a summary of information related t...

Страница 9: ...the device switches off or back on delivered therapy Stim Workspace Press the Stim tab to access the Stimulation Settings Workspace The Stimulation Settings Workspace is divided into five tabs which are used to Activate turn on up to four leads Adjust electrode configurations Measure impedance Set nominal values to begin stimulation Perform trial mapping Confirm the response and sensation of speci...

Страница 10: ...st be valid prior to activating the lead A valid lead configuration must include at least one positive and one negative electrode Lead Enable must be ON before the amplitude can be increased from 0 µA Location Enter the spinal level where stimulation therapy is delivered by this lead Pulse Parameters To select and change pulse parameters first press the desired increment level Fine Medium Coarse A...

Страница 11: ...to the programmed durations when out of session or in session in the Group Workspace this allows the clinician to evaluate and optimize On Off durations When in session in the Stim Workspace Periodic Stimulation will be inactive stimulation will remain ON to allow the clinician to optimize pulse parameters Rework Tabs Utility Rework This screen provides one button automation of the normal TNS rewo...

Страница 12: ...tient Programmer s Programmer Setup screen select the desired language then press the Bind Stimulator button Follow the sequence of prompts to enter the target NS serial number and complete the Binding process The Programmer will reboot upon completion Stimulator Unbinding On a bound Patient Programmer press the Unbind Stimulator button to remove the binding This will free the Patient Programmer t...

Страница 13: ... of the test level Mains power quality should be that of a typical commercial or home environment Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial hospital or home environment Guidance and Manufacturer s Declarations GUIDANCE AND MANUFACTURER S DECLARATION Electromagnetic Emi...

Страница 14: ...cy range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Recommended separation distances between portable and mobile RF communications equipment and the Spinal Modulation Neurostimulation System The Spinal Modulation Neurostimulation System is intended for use in an electromagne...

Страница 15: ... L C8 R C8 Magnet Turnoff Time Off 1 s 2 s 3 s 4 s 5 s 6 s 7 s 8 s 9 s 10 s 15 s 20 s 25 s 30 s 40 s 50 s 1 min 70 s 80 s 90 s 100 s 110 s 2 min Ramp Duration 1 s 2 s 3 s 4 s 5 s 6 s 7 s 8 s Appendix II Troubleshooting Pop up Messages Condition Buttons Resolution All stimulation has been turned OFF All stimulation turned off due to data corruption OK Contact your Spinal Modulation Representative C...

Страница 16: ...FF Stimulation can be turned off either by the All Stim OFF software button on the INS programmer or the Off switch on the TNS OK Stimulator battery has reached End of Service EOS Stimulation has been turned OFF permanently Battery has reached EOS voltage Stimulation is disabled in order to preserve power for RF communication OK Schedule replacement of the Stimulator Pop up Messages Cont Condition...

Отзывы: