USER  

MANUAL 

 

         BDM SERIES 

 

 

 

 

 

 

 

 

 

 

File: 2012-01-BDM-137 btA052 GB

Содержание BDM Series

Страница 1: ...USER MANUAL BDM SERIES File 2012 01 BDM 137 btA052 GB ...

Страница 2: ...5 5 Percent calculation functions PErC 21 15 6 Function for summing recipe ingredients rECIPE 22 15 7 Function for calibration with external weight CALIb 23 15 8 Label choosing function LAbEL 24 15 9 Function for setting serial interface parametr SErIAL 25 15 10 Printout configuration PrInt 26 15 11 Animal weighting function LOC 27 15 12 Tare memory function tArE 28 15 13 Maximum value indication ...

Страница 3: ...general use in stores in production control for sorting etc Scales comply with safety regulations of European Union what CE mark on name plate informs about and can be used for purposes where EC verification is not required NACE classification 33 20 31 2 Completeness Standard set consists of 1 Scale 2 Pan support 3 Pan 4 Feeder 5 Accumulators 6 pcs option 6 User manual 7 Guarantee card ...

Страница 4: ... type if feeder is supplied with the scale and supply voltage has to be compatible with specified technical data Do not use the scale when its cover is opened Do not use the scale in explosive conditions Do not use the scale in high humidity environment If the scale seems not to operate properly switch it off and do not use until checked by authorised service According to current acts of low about...

Страница 5: ...ons 250x180mm Working temperature 10 C 40 C Weighing time 3s Scale dimensions 255x310x107mm Scale weight 2 6kg Power supply 230V 50Hz 6VA 12V 1 2A external feeder Accumulators NI MH AA size 6 pcs Time of continuous work with accumulators 2200mAh c a 6 h with display backlight c a 16h without display backlight Time of automatic switching off the scale when working with accumulators 5 min AutoOFF fu...

Страница 6: ...__________________________ 5 General scale view General view 1 pan 2 pan support under pan 3 display 4 keys 5 water level 6 rotating legs Connectors view Zasilanie Zasilacz 230V RxD odbiór TxD nadawanie masa RS232C 1 2 3 4 5 6 7 8 9 6 5 1 2 4 3 Power supply 12V 500mA Feeder RxD read TxD transmit ground ...

Страница 7: ...ial function menu result printout HR increasing resolution of mass indication option indicator 0 zero indicator when scale platform is empty indicator of weighing result stabilisation NET net mass after use of T key MODE indicator of switching special function on bar indicator indicator of scale load 0 100 indicator OFF switching scale with key standby switch special function weighing pcs indicati...

Страница 8: ...ctromagnetic or magnetic fields 8 Preparing scale to work 1 Take the scale pan support pan and feeder out of the package It is recommended to keep the original scale package in order to transport the balance safely in future 2 Place the scale on a stable ground not affected by mechanical vibrations and airflows 3 Insert pan support 2 into holes in scale housing 4 Put pan 1 on pan support 5 Level t...

Страница 9: ...f a sample placed on the pan To make the control of a load on the pan easier and to avoid exceeding measurement range the scales with LCD display have load indicator calibrated 0 100 4 Weighing result should be read when the indicator lights which signalises result stabilisation 5 When the scale is not used but it is necessary for it to be ready to work it can be switched off by pressing I key The...

Страница 10: ...breaks in weighing Configuration of these mechanisms is done using b_LIGHT and Auto OFF functions 3 Charging accumulators is performed automatically after connecting feeder to the scale also during weighing Accumulator power level can be read using bAttErY function bat VoL option 11 Accumulators batteries replace 1 Open cover 11 Turning scale upside down take care that scale is not laid on the pan...

Страница 11: ...______________________________________________________________________________________________________ 2 Take accumulator container 12 off and put 6 pcs of AA accumulators in it The way of placing accumulators in container 12 13 ...

Страница 12: ...o work 13 Connection with computer or a printer The scale may send data to a computer or a printer via RS232C interface When cooperating with a computer the scale sends weighing result after initialising signal from a computer or after pressing key on the scale For cooperation with the scale a computer should be equipped with software allowing receiving data from the scale and using it further AXI...

Страница 13: ... scale sends 16 data according to the following format 16 Bytes transmission parameters 8 bits 1 stop no parity 4800 bps Description of particular bytes Byte 1 sign or space Byte 2 space Byte 3 4 digit or space Byte 5 9 digit decimal point or space Byte 10 digit Byte 11 space Byte 12 k l c p or space Byte 13 g b t c or Byte 14 space Byte 15 CR Byte 16 LF WK 1 connecting wire connects the scale wit...

Страница 14: ...oad on the pan take load off press a key when indication on the left is displayed forced change automatic change 14 1 Simple weighing If indication different from zero is shown for empty pan T key should be used Weighing results should be read when indicator is displayed 14 2 Weighing with tare The scale allows taring in the whole measurement range A sum of tare and net mass value can not exceed s...

Страница 15: ... function tArE maximum value indication function UP Force measuring function nEWton Anty disturbance filter option FILtEr Setting backlight function b_LIGHt Function for choosing reading unit rESOLUt Statistical calculations function StAt Basis weight of paper counting function PAPEr charging accumulators settings bAttErY accumulator option only display backlight settings b LIGht automatic switchi...

Страница 16: ...__________________________________________________ In further part of this manual the following graphical symbols will be used to describe scale functions put load on the pan take load off press the key when indication on the left is displayed forced change automatic change 12g T ...

Страница 17: ...g all available functions what can make operation time longer To distinguish ACtIV function from function menu list indicator is displayed on the left side In every moment it is possible to restore primary manufacture settings choosing dEFAULt special function The example on the left shows the operations of adding function for setting parameters of serial interface to function menu In order to rem...

Страница 18: ...tomatically maintained when there is no load on the pan or zero indication was received by pressing T key To switch the function on press MENU key and then using T key choose Autot and then Aut on To switch the function off press MENU key and then using T key choose Autot and AUt OFF Note During first 10 min after switching the scale on the function works automatically 0 01g AUtOtAr AUT AUt on AUt...

Страница 19: ...ion has the following options PCS OFF switch the function off PCS on switch the function on PCS recover last used unitary mass PCS 5 10 20 500 amount of pieces in a sample PCS SEt set any amount of pieces in a sample PCS MJ set unitary mass directly out leave menu without changes Comments 1 It is recommended that mass of single piece is greater than reading unit and mass of sample used in first ph...

Страница 20: ...ng weighing unit CarAt 1 ct 0 2 g carat MGrAM 1mg 0 001g milligram KGrAM 1kg 1000g kilogram Pound 1 lb 453 592374g English pound OunCE 1oz 28 349523g ounce OunCEt 1ozt 31 1034763g pharmaceutical ounce GrAIn 1gr 0 06479891g grain PennYW 1dwt 1 55517384g jewellery mass unit GrAM 1g gram The way of choosing carats as weighing unit is shown on the pictures on the left 1 60g UnIt ct CArAt 8 00 MENU T T...

Страница 21: ...e Weighing result is displayed in various formats depending on reference mass value For reference mass values between 0 3 5 of scale capacity format of weighing result is 100 0 for values between 3 5 35 it is 100 00 and above 35 100 000 The function has the following options PEr oFF switch the function off PEr on set current indication as 100 show indications in out exit without changes Caution 1 ...

Страница 22: ...C on start recipe weighing rEC Con continue previous recipe out exit without changes When proceeding with recipe successive ingredients A B C etc are weighed each time starting from zero indication which is obtained after scale taring If several ingredients are weighed their sum mass can be read despite several taring For this purpose use rEC oFF option Comments o indicator on the left side of sca...

Страница 23: ...acy should be used Operations sequence Press MENU key to display user functions shown one by one in loop Press T key when CALIbr function appears The following options will be displayed CAL oFF switch internal calibration off CAL on calibration with external weight out leave without changes Press T key when CAL on option appears Wait for writing zero to the scale When LOAD message appears put cali...

Страница 24: ...stant text Label patterns with number 4 digit used by user should be saved in scale memory according to printer manual Label pattern choice is made by inscribing label number using LAbEL function Press MENU button When LAbEL is displayed press T key Actual label number will show To enter new label number press T key to exit function without number change press MENU To inscribe label number use key...

Страница 25: ... serial interface SendInG stb transmission after key is used and result is stable transmission after key is pressed without need of stabilisation automatic transmission after load is put on and result is stable Auto continuous transmission about 10 results per second Cont Default parameter values Long 4800 bps 8 bits none stb In order to set needed parameters choose Port 1 function select appropri...

Страница 26: ...hoose this option to zero counter Id Prod product number 13 digits dAtE date option tIME time option Count counting result totAL result sum APW unitary mass netto net mass tArE current tare value brutto gross mass In order to make difference between Print function from menu list indicator is displayed on the left side Indicator o below informs which option ON or OFF is actually set In the pictures...

Страница 27: ...with key When the LOC on is displayed push the T key Tare the balance when necessary with T key and then put the animal on scale Wait until the weighting result is averaged the display will blink Next the balance will show final result and send it via serial port to a printer or computer The result is displayed for about 30 seconds In this time remove the animal from the pan Attention 1 The load s...

Страница 28: ...e pan is possible Writing tare vale using keys After MENU key is pressed and tArE function is chosen using T key the following options are displayed tAr OFF switch the function off tAr on switch the function on with previously used tare tAr call tare from memory tAr SEt write tare value to memory out leave the function Press T key when tAr SEt is displayed Pressing T key choose memory cell to whic...

Страница 29: ...memory cells will appear tAr 01 02 10 Cells with written values are marked with o indicator on the left and active value with indicator Choose appropriate cell using T key tArE function is activated with selected tare value From this moment the scale will display net mass i e a mass placed on the pan reduced by tare value Using T key causes zeroing the scale and then subtracting called tare value ...

Страница 30: ...ys zero indication before starting measurements tare Press MENU key Using T key choose UP function and then UP on Put on successively weighted objects scale display will show the maximum weight result Pressing the key will set the result to zero Attention Autozero function and the stabilisation indicator are deactivated when UP function is active The indication is the result of continuous averagin...

Страница 31: ...tal filter of chosen intensity during weighting The filter reduces influence of mechanical vibrations base vibrations air blast on measurement result Press MENU key and using T key choose FILtEr option The following options will show successively on the display FIL OFF filter off FIL on filter on Choosing option FIL on will display following intensity values of the filter After choosing intensity ...

Страница 32: ...acklight permanently on b_L ECO switch off after 30 seconds of inactivity no load changes and no key operation b_L bAt like above but when powering from accumulators only out leave without changes Switching backlight off causes decrease of energy consumption by the scale what is important during powering from accumulators 15 17 Function for choosing reading unit rESOLUt The function allows choosin...

Страница 33: ...r the scale is loaded and its indications stabilize After each loading printout is made with number of measurements result date and time if clock is installed and the function is activated For the obtained measurements series the scale evaluates n number of samples sum x sum of all samples n x x sum _ x average value sum x n min minimal value from n samples max maximal value from n samples max min...

Страница 34: ...e in Sta tAr automatic tare on off StA CFG function configuration Auto Automatic work samples are confirmed after loading the scale and indication stabilization ManuAL manual work confirmation is made by pressing key out exit from function Remember first to inscribe nominal weight value and tolerance mentioned above After that push T key when StA o is displayed Put on successively objects on the p...

Страница 35: ...all n samples AVERAGE average weight as Total n MIN minimum weight in n samples MAX maximum weight in n samples ST DEV standard deviation ST DEV standard deviation percentage To finish work with this function and zeroing result register press MENU key and then when StAt and Sta oFF is displayed press T button STATISTICS NOMINAL 50 000 g TOLERANCE 100 MAX N 500 NO SAMPLE TOL NOM TOL 1 10 007 g 2 20...

Страница 36: ... key in order to enter function menu While F PAP is displayed press T key While PAP n is displayed press T key Enter the quantity of paper slices using keys 0 digit increase T next digit MENU end Press T key when PAP ArE is displayed Enter surface of single paper slice in m2 as above Press T key when PAP on is displayed The scale will indicate paper substance g m2 sign on the right side In order t...

Страница 37: ... checking their power level The function has the following options bAt OFF charging off option required if ordinary batteries are used bAt on charging on accumulators are being charged even after switching scale off using I key bAt VoL reading power level of accumulators in go back to mass indication pressing MENU key out leave without changes An attempt of charging ordinary batteries can cause se...

Страница 38: ...cale off automatically during inactivity what allows decreasing effective energy consumption and making time of working with accumulators longer AOF OFF scale is not switching off AOF on scale is switching off after 5 minutes of inactivity no load changes and no key operation AOF bAt like above but when powering from accumulators only out leave function without changes 0 00g Auto OFF AOFon 0 00g 5...

Страница 39: ...earing the adding register tot report printout after each measurement tot report printout disabled Press the T key when tot is displayed Place successively samples on the pan and press after each measurement to store the value in the adding register The storing is confirmed by on the display To display current results enter totAL function menu and choose tot Prn option The results are display in t...

Страница 40: ...nt functions Report printout sample Attention Maximum number of measurements 99 999 Maximum total load 99 999 000d The weighing unit of the aggregated value from the register total is the same as the weighing unit stated on the keypad or is 1000 times greater signalised with o indicator at the left of the display If the registered value is too big to be displayed E communicate appears on the displ...

Страница 41: ...e actual date and time dAt PIn acces code insertion dAt For date print format selection EU or USA The example at the left presents how to use dAt SEt option In order to write actual time and date T key should be pressed while dAt Set is displayed Confirm numbers displayed successively using T key to get the right time and date 0 key pressed repeatedly enables speeding up the process Time format h ...

Страница 42: ...d greater than upper threshold the balance displays MAX with the sound signal when exceeding the threshold If the balance is equipped with THRESHOLDS control connection the comparison result may be used to control optical signalling devices SIGNAL mode batching devices IMPULS mode In standard configuration the scale is set up to cooperate with optical signalling device On the outputs P1 and P2 the...

Страница 43: ... start the function with the excess signalisation SEt 1 set lower threshold value SEt 2 set upper threshold value SEt 3 set zero excess signalisation Using T key choose option SEt 1 Set the lower and the upper threshold values with the following keys digit increase decimal point T next digit MENU end Next choose option SEt 2 and write upper threshold value When necessary use SEt 3 option to set th...

Страница 44: ...des e g 1N4148 The balance producer offers ready electronic PCB MS3K P which contains RM96P relays with input voltage DC24V and output AC 250V 3A Important notes 1 After switching the balance on both thresholds are set as maximum values 2 When selecting upper threshold value pay attention that its value is not lower than the lower threshold value 3 Selecting lower and upper threshold value is also...

Страница 45: ...ee card and on website www axis pl 6 Scales can be sent for repair as messenger delivery only in original package if not there is a risk of damaging the scale and loosing guarantee Failure messages Message Possible cause Recommendation C 1 6 more than 1min negative result in one of autotests contact service centre if the message remains scale is not weighing protecting screw remains in the scale r...

Страница 46: ...ment of radio disturbance characteristics of information technology equipment and IEC 61000 4 3 Electromagnetic compatibility EMC Part 4 3 Testing and measurement techniques Radiated radio frequency electromagnetic field immunity test harmonized with the Council Directive 89 336 EEC Additional information Conformity evaluation for the Council Directive 89 336 EEC were carried out by Laboratorium B...