background image

b. AVG: measuring average temperature
c. MIN: measuring minimum temperature
d. DIF: Basic on the set DIF value, compute the difference of current reading.
2Press the “MODE” key for 3 seconds, then press it for a short time, LCD will flickering display ε-LAL-HAL;
(This item have memory function, it will show the same measure mode at the next time you open this item)
a.ε:emissivity can use �/� key to set from 0.10 to 1.00, press “MODE” key to confirm.
b. LAL: low temperature alarm--when selected LAL, short press �/� key to set the temperature alarm slowly
and long press �/� key to set the temperature alarm quickly, confirmed by long pressing “MODE” key; When
the tested temperature is less than the alarm which you set, LCD display icon with the buzzer alarm.
c. HAL: high temperature alarm--when selected HAL, short press �/� key to set the temperature alarm
slowly and long press �/� key to set the temperature alarm quickly, confirmed by long pressing “MODE”
key; When the tested temperature is more than the alarm which you set, LCD display icon with the buzzer
alarm.
(7) Display area
(8) Laser emission area
(9) Sticker
(10) Battery cover: you need to open it to replace the battery.

9. Specification

Accuracy: ±(a% reading)
Working environment temperature: 0℃~40℃ (32°F~104°F) can not guarantee the accuracy
Store temperature: -20℃~50℃ (-4°F~122°F)
Relative humidity: 10-95% RH non-condensed
Weight/Size: 270g (include battery); 141*200*60mm
Power: 9V battery (not included)
Temperature range ///// -50℃~1150℃ (-58°F~2102°F)
Distance ratio ///// 50:1
Emissivity ///// 0.10~1.00 adjustable(preset 0.95)
Accuracy ///// - 50℃~0℃ (- 58°F~32°F) - ±3℃/±5°F; (take the bigger value) 0℃~100℃ (32°F~212°F) -
±1.5℃/±2.7°F; 100℃~1150℃ (212°F~2102°F) - ±1.5%
Resolution ///// 0.1℃/°F
Response time ///// <250ms
Dual range display ///// √
MAX/MIN value ///// √
D-value / Average value ///// √
High/Low temperature alarm ///// √
℃/°F setting ///// √

Содержание AX-7540

Страница 1: ...m out and check below accessories 1 Operation manual 1pc If you found that there s anything missing or damage please contact your supplier 3 Warning Cautions 3 1 Warning To avoid the potential situati...

Страница 2: ...r the target the closer measure distance When accuracy is critical make sure the target is at least twice as large as the spot size 5 Emissivity Most organic materials and painted or oxidized surfaces...

Страница 3: ...er 0 93 Wood 0 94 6 Operation 1 Fast measurement 1 Open the battery door and insert a 9V battery properly 2 Pull the trigger to turn on the unit 3 Aim at the target surface with the laser and pull the...

Страница 4: ...7 LCD display 1 LCD display a measuring reading b measuring unit c laser on icon d back light on icon...

Страница 5: ...Press the trigger again you can continue to measure The data will be hold for about 20 seconds and the item will auto power off in 20 seconds if without any action 2 Back light button Pull the trigger...

Страница 6: ...d HAL short press key to set the temperature alarm slowly and long press key to set the temperature alarm quickly confirmed by long pressing MODE key When the tested temperature is more than the alarm...

Страница 7: ...eaning Blow off loose particles using clean compressed air Gently brush remaining debris away with a moist cotton swab The swab may be moistened with water Case cleaning Clean the case with a damp spo...

Отзывы: