17
Test Condition: Forward DC current approximately 1mA. Reversed DC voltage approximately 2.8V
4.2.6.
Capacitance
Range
Resolution
Accuracy
6nF
1pF
±(5,0% +5)
60nF
10pF
±(3,0% +3)
600nF
100pF
6
µ
F
1nF
60
µ
F
10nF
±(5,0% +3)
600
µ
F
100nF
6mF
1
µ
F
60mF
10
µ
F
uncertainty
4.2.7.
Current
DCA:
Range
Resolution
Accuracy
600
µ
A
0.1
µ
A
±(1.0% +3)
6000
µ
A
1
µ
A
60mA
0.01mA
±(1.5% +3)
600mA
0.1mA
10A
10mA
±(1.8% +5)
ACA:
Range
Resolution
Accuracy
600
µ
A
0.1
µ
A
±(1.5% +5)
6000
µ
A
1
µ
A
60mA
0.01mA
±(1.8% +8)
600mA
0.1mA
10A
10mA
±(2% +8)
Above accuracies can be guaranteed within 5%
~
100% of the full range.
The true RMS meter has residual value within 10 count when the test leads are shorten, but that
will not affect the accuracy of measurement.
Overload protection:
F 10A/1000V fuse for 10A range
F 0.63A/1000V fuse for
µ
A and mA ranges
Maximum input current:
600mA dc or 600mA ac rms for
µ
A and mA ranges
10A dc or 10A ac rms for 10 A ranges.
For measurements >6A, 4 minutes maximum ON to measure 10 minutes OFF; above 10A unspecified.