background image

4. Resistance measurement

5. Safety Notices

• When voltage measurement selection button does not be pressed, it is possible to appear the high voltage at

the output voltage terminal.

• When measuring, first check the measurement voltage selection and if the indication of the measurement

voltage on LCD is in accordance with the voltage needed

• To ensure the safety of operation, the operation cannot be done until the measured object removed from the

power supplied by the electricity grid system, and should be discharged fully to testify that the measured
object does not have any danger of power.

• When measuring, it is not allowed to handle the terminal of measurement to make sure the accuracy of reading

and personal safety.

• Keep the instrument away from the high temperature place, and from the direct sunshine to avoid affecting

the life of the LCD.

• When low battery symbol of “BATTERY” appears, please replace the batteries in time. Take the batteries

out when storage to avoid the leakage of batteries to damage the instrument.

• When unload, reading is normal, it won’t affect the measurement.
• When measuring MΩ, if the reading is unstable, it is likely caused by the interference of surrounding and

unsteadiness of the insulated materials, it will make the reading stable to connect the “G” terminal to the
shield side of the measured object

Содержание AX-T2200

Страница 1: ...LCD display Display the measured data and units 10 Instrument model 11 L Terminal for connecting the measured circuit 12 G Terminal for protection connect the electrode wire of protection loop to G terminal when the measured object is required to add the protection loop to eliminate the leakage effect 13 ACV Input terminal of AC Voltage measurement 14 E Terminal for connecting the ground of the me...

Страница 2: ...peration environment Temperature 0 C 40 C Relative humidity 30 RH 85RH Dimension 175 L 110 W 70 D mm Weight 630g including batteries 2 2 Technical property Measured voltage 250V 10 500V 10 1000V 10 RANGE 0 1MΩ 20MΩ 0 1MΩ 50MΩ 0 1MΩ 100MΩ 20MΩ 500MΩ 50MΩ 1000MΩ 100MΩ 2000MΩ Accuracy 4 of reading 2 digits Shot current 1 8mA 1 8mA 1 6mA Median resistance 2 MΩ 2MΩ 5MΩ Terminal position L E ACV750V Acc...

Страница 3: ...rement Connect the electrode of the measured object to the relative terminal of the instrument When measuring cable connect G Terminal to the protection loop Press the button of measurement and start it turn the knob right to lock the button when the value is stable get the reading from LCD Connect the input line E to the ground of the measured object and L to the circuit of the measured object re...

Страница 4: ...ot have any danger of power When measuring it is not allowed to handle the terminal of measurement to make sure the accuracy of reading and personal safety Keep the instrument away from the high temperature place and from the direct sunshine to avoid affecting the life of the LCD When low battery symbol of BATTERY appears please replace the batteries in time Take the batteries out when storage to ...

Страница 5: ...e do not change the leads at will When the external AC adaptor supplies the power the inside batteries are disconnected and the batteries can t be charged at the time Note please select power supply method 6 Sets of instrument Digital Insulation Tester 1pc 10A test leads 1set Operation Manual 1copy Shoulder strap 1pc ...

Отзывы: