15
4.2.1 Voltage
DCV:
Range
Resolution
Accuracy
600mV
0.1mV
±(0.5% of rdg +8 digits)
6V
1mV
±(0.8% of rdg +5 digits)
60V
10mV
600V
100mV
1000V
1V
±(1.0% of rdg +10 digits)
ACV:
Range
Resolution
Accuracy
600mV
0.1mV
±(3.0% of rdg + 3 digits)
Manual Range only
6V
1mV
±(1.0% of rdg + 3 digits)
60V
10mV
600V
100mV
700V
1V
±(1.5% of rdg + 3 digits)
1) Frequency Range for ACV: 40Hz~400Hz.
2) Response for ACV: Average, calibrated in rms of sine wave.
3) Overload Protection: 1000V dc or 700V ac rms.
4) Input Impedance (Nominal): DC voltage: >10M
Ω
; AC voltage: >10M
Ω
5) Common Mode Rejection Ratio: DC voltage: >100dB at dc, 50 or 60 Hz; AC voltage: >60dB at dc,
50 or 60 Hz
6) Normal Mode Rejection Ratio: DC voltage: >45dB at 50 or 60Hz
4.2.2 Frequency (10Hz-1MHz)
Range
Resolution
Accuracy
99.99Hz
0.01 Hz
±(0.1% of rdg+3digits)
999.9Hz
0.1 Hz
9.999kHz
0.001kHz
99.99kHz
0.01kHz
999.9kHz
0.1kHz
Overload Protection: 250V dc or 250V ac rms
Содержание AX-18B
Страница 1: ...AX 18B INSTRUCTION...