The calibration card is supplied with the PV-1072 and is kept in the wallet.
With the
Ready to Scan
message showing on the screen,
Press the
Menu/select
key.
And the screen will show
►
Information
Press the
Scroll down
button until the screen displays
►
Calibrate
Press the
Menu/Select
Button
There will be a
♦
Set RMin
message with a
5
below. Look at your calibration card to
find the value for RMin.
Press
Menu/Select
and you can adjust the figure shown (if necessary) by using the
Scroll up
and
Scroll down
buttons. When the numbers match, press
Menu/Select
to
confirm the setting. Press the
Scroll down
button again and the screen will display
♦
Set RMax
message with a
85
below.
Look at your calibration card to find the value for RMax. Press
Menu/Select
and
you can change the value by using the
Scroll up
and
Scroll down
buttons until it
matches the value shown on the calibration card. Press
Menu/Select
to confirm the
setting. Press the
Scroll down
button and the message will be displayed
♦
Calibrate
Press the
Menu/ Select
button.
The message will be displayed
Calibrate
Cal. Scan (0/3)
Pick up the verifier read-head, place it over the barcode on the calibration card in
such a way that all four corners of the read-head are touching the card and then
press the trigger on the read-head. The message will change to read
Cal. Scan
(1/3)
. Press the trigger two more times and the display will first read
Cal. Scan
(2/3)
and then
Calibration Successful !
Press the
Clear/Back
button twice to see the message
Axicon Verifier
Ready to scan
Calibration