SC-Series Manual

Model

Acceptable 

Tolerance (±)

Calibration 

Weight

Recommended 

Accuracy Class

SC-501,  

SC-501A

0.02g

200g,  

400g

OIML M2

SC-2KG, 

SC-2KGA

0.2g

1000g, 

2000g

OIML M2

VIII.  Adjustment 

Do Not Proceed Without Proper Weights (see above)

1.  Turn the scale ON.
2.  Press and hold the 

MODE

 key until the display 

shows “CAL”, then release.

3.  Make sure no weight is on the platform and press 

MODE 

again. The display will flash “CAL” followed 

by the first required calibration weight. 

4. 

Place the first required weight on the platform.

5. 

Once the first weight is confirmed, the display will 

flash the second required calibration weight. Place 

the required weight on the platform.

6. 

Once the second weight is confirmed, the display 

will show “PASS” and return to normal weighing 

mode. Calibration is complete!

IX. Troubleshooting

• 

 

Problem:

 Scale will not turn on

• 

 

Solution: 

Check the batteries and battery contacts.

• 

 

Problem: 

Displayed weight fluctuates randomly

• 

 

Solution: 

Make sure your workspace is stable and    

Содержание SC-501

Страница 1: ...American Weigh Scales SC 501 SC 501A 500g x 0 01g SC 2KG SC 2KGA 2000g x 0 1g User Manual SC Series Copyright 2011 American Weigh Scales Inc All rights reserved Rev 2 0...

Страница 2: ...30 60 seconds before performing calibration Store your scale in a clean dry location Dust dirt and moisture can accumulate on the weighing sensors and electronics causing inaccuracy or malfunction Avo...

Страница 3: ...AA Batteries x2 4 AC Adapter II Key Description 1 ON OFF key Press to turn the scale ON or OFF 2 MODE key Changes the displayed weighing unit 3 TARE key Re zeros the scale Used to remove the weight of...

Страница 4: ...y use the included adapter when plugging the scale into AC power SC 501A SC 2KGA only IV Operation 1 Place the scale on a flat level surface and press the ON OFF key to turn the scale on 2 Once the st...

Страница 5: ...container with the quantity to be weighed The display will show the net weight of the contents 4 To return to normal weight display gross weight remove any items from the weighing platform and press t...

Страница 6: ...The display will flash CAL followed by the first required calibration weight 4 Place the first required weight on the platform 5 Once the first weight is confirmed the display will flash the second r...

Страница 7: ...stment must be performed with the weights listed in the chart on the previous page or inaccuracy will occur X Error Codes 1 OUTZ or OUE2 Zero point has drifted out of the acceptable range Try to perfo...

Страница 8: ...g 0 3g Power 2 x 1 5V AAA Alkaline or AC Adapter 5V 500mA Auto Off Time 90 Seconds of Non Use Disabled when using AC Adapter Dimensions 4 1 x 5 0 x 0 7 in Platform Dimensions 4 0 x 4 0 in Warranty 10...

Отзывы: