SC-Series Manual
Model
Acceptable
Tolerance (±)
Calibration
Weight
Recommended
Accuracy Class
SC-501,
SC-501A
0.02g
200g,
400g
OIML M2
SC-2KG,
SC-2KGA
0.2g
1000g,
2000g
OIML M2
VIII. Adjustment
Do Not Proceed Without Proper Weights (see above)
1. Turn the scale ON.
2. Press and hold the
MODE
key until the display
shows “CAL”, then release.
3. Make sure no weight is on the platform and press
MODE
again. The display will flash “CAL” followed
by the first required calibration weight.
4.
Place the first required weight on the platform.
5.
Once the first weight is confirmed, the display will
flash the second required calibration weight. Place
the required weight on the platform.
6.
Once the second weight is confirmed, the display
will show “PASS” and return to normal weighing
mode. Calibration is complete!
IX. Troubleshooting
•
Problem:
Scale will not turn on
•
Solution:
Check the batteries and battery contacts.
•
Problem:
Displayed weight fluctuates randomly
•
Solution:
Make sure your workspace is stable and