AWS QS-500 Скачать руководство пользователя страница 5

4.  To return to normal weight display (gross weight), 

remove any items from the weighing platform and 

press the

 ON/OFF/TARE

 key again to re-zero the 

scale.

VII.  Calibration

Calibration and adjustment are performed at the factory. 

However, the weighing range can shift slightly during 

shipping. Also, due to changes in local acceleration of gravity 

around the world, adjustment is recommended when moving 

the scale from one location to another. To check calibration:

1.  Place the required calibration weight on the scale 

and note the displayed value.

2.  If the displayed value is not within acceptable 

tolerance (see chart below), adjustment is required.

Model

Acceptable 

Tolerance (±)

Calibration 

Weight

Recommended 

Accuracy Class

QS-500

0.2g

500g

OIML M2

VIII.  Adjustment

1.  With the scale OFF, press and hold the 

MODE

 key 

then press the 

ON/OFF/TARE key

 until the display 

shows a random number. Now release both keys. 

2.  Press the 

ON/OFF/TARE

 key again. The display will 

show “Zero” then “500.0g”. 

3.  Gently place a 500g weight on the center of the 

platform and press the 

ON/OFF/TARE

 key again. 

The LCD will display “PASS” then “500.0”. 

4.  Calibration is complete. If the scale shows “ H ”, an 

error has occurred - Turn off the scale and repeat 

steps 1 to 3 above.

Отзывы: