background image

remove any items from the weighing platform and 

press the

  /T

 key again to re-zero the scale.

VII.  Calibration

Calibration and adjustment are performed at the factory. 

However, the weighing range can shift slightly during 

shipping. Also, due to changes in local acceleration of gravity 

around the world, adjustment is recommended when moving 

the scale from one location to another. To check calibration:

1.  Place the required calibration weight on the scale 

and note the displayed value.

2.  If the displayed value is not within acceptable 

tolerance (see chart below), adjustment is required.

Model

Acceptable 

Tolerance (±)

Calibration 

Weight

Recommended 

Accuracy Class

PV-650

0.2g

500g

OIML M2

VIII.  Adjustment

1.  Turn the scale OFF.
2.  Press and hold the 

M

 and 

/T

  keys together until 

the display shows “CAL”.

3.  Press the 

/T

 key again. The display will flash 

“Zero” followed by the calibration weight value (see 

chart).

4.  Place the calibration weight on the center of the 

weighing platform and press 

/T 

once more. 

5.  The display will show “CAL” then “PASS” and return 

to normal operation. 

 

Отзывы: