negative number.

3. 

You may now fill the container with the quantity to be weighed. 

The display will show the net weight of the contents.

4. To return to normal weight display (gross weight), remove any 

items from the weighing platform and press the

 TARE

 key 

again to re-zero the scale.

VIII. Calibration (Advanced Users Only)
Calibration and adjustment are performed at the factory. Only 

trained technicians should perform calibration and adjustment 

if the scale is not weighing properly. Do not proceed with 

calibration without the proper 10kg weight with class M2 

or better or you could cause your scale to be inaccurate. 

Calibration is not covered under warranty.

1. Place the required calibration weight on the scale and note the 

displayed value.

2. If the displayed value is not within acceptable tolerance (see 

chart below), adjustment is required. 

Model

Acceptable 

Tolerance (±)

Calibration 

Weight

Recommended 

Accuracy Class

PS-25

4g

10kg

OIML M2

 

IX.  Adjustment (Advanced Users Only)

1. Press and hold the 

MODE

 key until the display shows “CAL”.

2. Press the 

MODE

 key again. The display will flash “CAL” 

followed by the calibration weight value (see chart).

3. Place the required calibration weight on the center of the 

platform. 

4. The display will show “PASS” then return to normal weighing 