AWS MSC-200 Скачать руководство пользователя страница 6

Mouse Scale

the weight and press 

 

to turn the scale off. 

Calibration is complete!

X.  MSC-650 Adjustment

MSC-200 Calibration Adjustment

1.  Turn the scale OFF.
2.  While holding the [ M ] key, tap the 

 

key to turn 

the scale on then release both keys.

3.  The display will show a random number, press the 

 

key again.

4. 

The display will show “Zero” followed “500.0”. Place 

the required weight on the platform then press the 

 

key again.

5. 

The display will show “PASS”. Remove the weight 

and press 

 

to turn the scale off. Calibration is 

complete!

XI. Troubleshooting

• 

 

Problem:

 Scale will not turn on

• 

 

Solution: 

Check the batteries

• 

 

Problem: 

Displayed weight fluctuates randomly

• 

 

Solution: 

Make sure your workspace is stable and    

Содержание MSC-200

Страница 1: ...User Manual Copyright 2010 American Weigh Scales Inc All rights reserved Rev 3 0 MSC 200 200g x 0 01g MSC 650 650g x 0 1g Mouse Scale...

Страница 2: ...ation and adjustment if available Store your scale in a clean dry location Dust dirt and moisture can accumulate on the weighing sensors and electronics causing inaccuracy or malfunction Avoid using y...

Страница 3: ...are 1 Plug the small end of the included USB cable into the Mouse Scale 2 Insert the large end of the USB cable into an open USB port on your computer 3 The computer should recognize the new input dev...

Страница 4: ...e is turned off then back on it will default to the last unit used MSC 200 Units g gram oz ounce ct carat gn grain MSC 650 Units g gram oz ounce ozt troy ounce pennyweight dwt VII Using the Tare Featu...

Страница 5: ...check calibration 1 Place the required calibration weight on the scale and note the displayed value 2 If the displayed value is not within acceptable tolerance see chart below adjustment is required...

Страница 6: ...display will show a random number press the key again 4 The display will show Zero followed 500 0 Place the required weight on the platform then press the key again 5 The display will show PASS Remove...

Страница 7: ...aximum capacity exceeded Remove the extra weight to avoid damaging the load cell 3 L Batteries are low Replace the batteries MSC 200 MSC 650 Max Capacity 200g 650g Readability 0 01g 0 1g Min Capacity...

Отзывы: