Calibration

Calibration

 (after warming up)

Action

Description

Press [ ON/

OFF ]

1.  Power off the scale

Press and 

Hold [UNIT] 

then press 

[ON/OFF]

2.  While holding the UNIT key, tap the ON/OFF key to turn the scale 

on. The display will show CALu=. The scale is now in calibration 

mode.

Choose Unit 3.  Press the UNIT key to toggle between weighing units (kg, g, lb, oz).

Press [ZERO] 4.  Press ZERO to confirm the weighing unit and proceed to 

calibration.

Enter the  

Mass

5.  Using the UNIT and TARE keys, enter the mass value of the 

calibration weight you are going to use:

 

Press the UNIT key to toggle between digits

 

Press the TARE key to increment the number

 

I

note: 

The ideal calibration weight is between 2/3 capacity to full 

capacity in order to get the most accurate results.

Press [ZERO] 6.  Press ZERO to confirm the mass value. The display will show a 

random number (the AD value).

Press [UNIT] 7.  Make sure there is nothing on the scale then press UNIT to 

confirm.

8.  When the display flashes the calibration mass value, place the 

required mass on the platform. 

Press [UNIT] 9.  Once the weight stabilizes, press the UNIT key to confirm the 

calibration adjustment. The display will return to the AD value

10.  Remove the calibration weight and turn the scale off. Calibration 

is complete!

 

I

note: 

Calibration should be done using accurate weights. The scale will 

accept the weight you enter, even if it is not correct. Take care to enter the 
calibration mass value correctly and only use accurate weights.

Содержание HX-3001

Страница 1: ...HX 502 500g x 0 01g HX 3001 3000g x 0 1g HX Series Copyright 2014 American Weigh Scales Inc All rights reserved Rev 1 0...

Страница 2: ...2 Scale Assembly 2 Key Explanation 3 Display Explanation 3 Getting Started 4 Basic Weighing 4 Calibration Adjustment 5 Function Setting Mode 6 Back light Setting 6 Auto Off Setting 6 Returning to Norm...

Страница 3: ...dverse effect on the scale s accuracy I I The scale should not be used in hazard areas that have an explosion risk Service Use only authorized parts and accessories We are not responsible for any modi...

Страница 4: ...ght which may cause it to heat up Large temperature fluctuations may require recalibration of the scale I I Avoid dust dirt and moisture which may accumulate on the weighing sensors and cause a malfun...

Страница 5: ...TARE ZERO Remove any items from the weighing platform then press to Zero the scale Tare Key ON OFF ON OFF UNIT TARE ZERO Tares the scale Use this to subtract the weight of an empty container 6 Display...

Страница 6: ...le 2 If using a container or tray place it on the platform example 20g Press TARE 3 Press the Tare key to subtract the weight of the container or tray 4 Place the sample in the container example 100g...

Страница 7: ...I I Note The ideal calibration weight is between 2 3 capacity to full capacity in order to get the most accurate results Press ZERO 6 Press ZERO to confirm the mass value The display will show a rando...

Страница 8: ...t off inactive Press ZERO 2 Press ZERO to toggle between A ON Auto Shut off active or A OFF Auto Shut off inactive Backlight Settings Action Description Press UNIT 1 While still in Function Setting Mo...

Страница 9: ...l Err b Low battery Replace the batteries Err 2 Platform is not secure Check to make sure that the platform is installed correctly Err S Inner code unstable Err C Calibration weight value has not been...

Страница 10: ...60 Specifications Model Capacity Readability Repeat ability Linearity Oper Temp Pan Size Overall Dimensions HX 502 500g 0 01g 2d 2d 32 104 F 5 7 x 5 7 in 2 7 x 7 0 x 9 0 in HX 3001 3000g 0 1g 1d 2d 32...

Отзывы: