6. Remove the sample quantity from the platform.

7. You now can place any amount of your item(s) on 

the scale and the LCD will display a count.

8. You can switch between weighing mode and 

counting mode by pressing [ 

MODE

 ] or [ 

PCS

 ] key 

respectively.

IV. CALIBRATION

Calibration is NOT required before first use. Only

calibrate if the scale is not weighing properly.

The CG-500 calibrates with a 500g weight*

The CG-50 calibrates with a 50g weight*

*Weights not included.

1. Turn on the scale by pressing the [ 

ON/OFF

 ] key.

2. Wait until the LCD displays “0.0” or "0.00", then 

press and HOLD the [ 

MODE

 ] key for 3 seconds. The 

LCD will display “CAL”. You may now release the

MODE

 ] key. 

3. Press the [ 

MODE

 ] key again. The LCD will display 

“CAL” followed by the required calibration weight (see 

above).

4. Place the required calibration weight on the center 

of the platform. The display will show “PASS” and 

return to normal weighing mode.

5. Put away the calibration weight and press

ON/OFF

 ] to turn off the scale and the calibration is 

finished!