background image

VI. Using the Tare Feature

You can use this scale’s tare feature to subtract the weight 

of an empty container or vessel for net weight determination. 

NOTE: The weight of the container will subtract from the 

available capacity (subtractive taring). 

1.  Place the empty container or vessel on the weighing 

platform.

2.  Press the 

T

 key once. The display will then reset to 

zero. If the container is removed, its weight will be 

displayed as a negative number.

3. 

You may now fill the container with the quantity to 

be weighed. The display will show the net weight of 

the contents.

4.  To return to normal weight display (gross weight), 

remove any items from the weighing platform and 

press the

 T

 key again to re-zero the scale.

VII.  Calibration

Initial calibration is performed at the factory. However, the 

calibration can drift slightly due to stresses from shipping. 

Also, due to changes in local acceleration of gravity around 

the world, adjustment is recommended when moving the 

scale from one location to another. To check calibration:

1.  Place the required calibration weight on the scale 

and note the displayed value.

2.  If the displayed value is not within acceptable 

tolerance (see chart on next page), adjustment is 

required.

Содержание AERO-100

Страница 1: ...American Weigh Scales AERO 100 100g x 0 01g AERO 650 650g x 0 01g User Manual AERO Copyright 2011 American Weigh Scales Inc All rights reserved Rev 1 0...

Страница 2: ...60 seconds before performing calibration if available Store your scale in a clean dry location Dust dirt and moisture can accumulate on the weighing sensors and electronics causing inaccuracy or malfu...

Страница 3: ...1 Mode Key Changes the displayed unit of measure 2 Tare Key Re zeros the scale Used to remove the weight of a tray or container 3 Calibration Key Enters calibration adjustment mode 4 Power Key Turns t...

Страница 4: ...tive contacts are properly aligned IV Operation 1 Place the scale on a flat level surface and press the key to turn the scale on 2 Once the start up process completes and the scale goes to zero you ar...

Страница 5: ...e weighed The display will show the net weight of the contents 4 To return to normal weight display gross weight remove any items from the weighing platform and press the T key again to re zero the sc...

Страница 6: ...of the scale platform The display will flash PASS then return to normal weighing Adjustment is complete NOTE Adjustment using the incorrect mass will result in inaccurate readings IX Troubleshooting...

Страница 7: ...Lo Batteries are low Replace the batteries XI Specifications AERO 100 AERO 650 Max Capacity 100g 600g Readability 0 01g 0 1g Min Weight 0 03g 0 2g Power 2 x 1 5V AAA Alkaline Dimensions 0 7 x 2 6 x 4...

Страница 8: ...AERO Manual American Weigh Scales Inc Norcross GA 30092 Made In China to our exact specifications and quality control...

Отзывы: