Preparing the Gas Analysis System for Use
Note
the information given is a recommendation and is based on well-founded
series of measurements. But there is always the possibility of changed circumstances, which might
require other factors. Therefore Awite shall not be liable for any deviations of the flow profiles.
Figure 10: Inlet and Outlet Sections Flow Sensor AwiFLOW
Table 2: Recommendations for Installing an AwiFLOW Flow Sensor
Standard Recommendation for Straight Pipeline with Undefined Pipe Routing
Pipe inner diameter
Inlet section L1
Outlet section L2
Illustration
< 150mm
>= 150mm
with flow conditioner
20 x pipe Ø
15 x pipe Ø
6 x pipe Ø
10 x pipe Ø
7.5 x pipe Ø
2 x pipe Ø
(A)
Recommendation for 45° or 90° angle on one level
< 150mm
>= 150mm
with flow conditioner
15 x pipe Ø
10 x pipe Ø
6 x pipe Ø
7.5 x pipe Ø
5 x pipe Ø
1 x pipe Ø
(B)
Recommendation for 45° or 90° angle on two levels
< 150mm
>= 150mm
with flow conditioner
20 x pipe Ø
15 x pipe Ø
6 x pipe Ø
10 x pipe Ø
5 x pipe Ø
2 x pipe Ø
(C)
Recommendation for a Flow Control Valve in the Inlet or Outlet Section
< 150mm
>= 150mm
20 x pipe Ø
15 x pipe Ø
10 x pipe Ø
5 x pipe Ø
(D)
31