MECHANICAL INFORMATION

TOP COVER REMOVAL

If necessary, the interior of the instrument may be accessed by removing the four 
Phillips screws on the top panel. With the four screws removed, the top cover may be 
slid back (and off).

 Always disconnect the power cord and allow the instrument to sit unpowered for 10 

minutes before opening the instrument. This will allow any internal stored charge to 
discharge.

There are no user-adjustable internal circuits. For repairs other than fuse replacement, 
please contact Avtech ([email protected]) to arrange for the instrument to be 
returned to the factory for repair. Service is to be performed solely by qualified service 
personnel.

 Caution: High voltages are present inside the instrument during normal operation. 

Do not operate the instrument with the cover removed.

RACK MOUNTING

A rack mounting kit is available. The -R5 rack mount kit may be installed after first 
removing the one Phillips screw on the side panel adjacent to the front handle.

ELECTROMAGNETIC INTERFERENCE

To prevent electromagnetic interference with other equipment, all used outputs should 
be connected to shielded loads using shielded coaxial cables. Unused outputs should 
be terminated with shielded coaxial terminators or with shielded coaxial dust caps, to 
prevent unintentional electromagnetic radiation. All cords and cables should be less 
than 3m in length.

32

Содержание AVR-E1-B

Страница 1: ...MBER ____________ A V T E C H E L E C T R O S Y S T E M S L T D N A N O S E C O N D W A V E F O R M E L E C T R O N I C S S I N C E 1 9 7 5 info avtechpulse com http www avtechpulse com Tel 888 670 87...

Страница 2: ...ace said defective item This warranty does not apply to units which have been dissembled modified or subjected to conditions exceeding the applicable specifications or ratings This warranty is the ext...

Страница 3: ...E LICENSING 10 INSTALLATION 11 VISUAL CHECK 11 POWER RATINGS 11 CONNECTION TO THE POWER SUPPLY 11 PROTECTION FROM ELECTRIC SHOCK 12 ENVIRONMENTAL CONDITIONS 13 LABVIEW DRIVERS 13 FUSES 14 AC FUSE REPL...

Страница 4: ...34 WIRING OF AC POWER 34 PCB 158R4 LOW VOLTAGE POWER SUPPLY 35 PCB 235B HIGH VOLTAGE DC POWER SUPPLY 36 PCB 151D RANGE POLARITY RELAY BOARD 37 PCB 217A RELAY DRIVER 38 PCB 96F HIGH VOLTAGE DC POWER SU...

Страница 5: ...s with the PN suffix can generate both polarities A 50 Ohm load is required for proper operation The output stage may be damaged if the output is not terminated into a 50 load The AVR E1 B is a highly...

Страница 6: ...E1 B is available with several options including OS Option an externally generated DC offset can be added to the output R5 Option This is the optional rack mounting kit The R5 rack mount kit may also...

Страница 7: ...fter 10 minute warm up Amplitude Typically 3 of setting 2 of maximum Delay Period Typically 3 of setting 5 ns Pulse width Typically 3 of setting 2 ns at maximum amplitude As the amplitude is reduced t...

Страница 8: ...g Also check that your load is adequately shielded It may be necessary to enclose the load in a metal enclosure If any of the connectors on the instrument are unused they should be covered with shield...

Страница 9: ...under the sole responsibility of Avtech Electrosystems Ltd Specifically products manufactured do not contain the substances listed in the table below in concentrations greater than the listed maximum...

Страница 10: ...ecycling activities to local governments and financing of recycling activities FIRMWARE LICENSING Instruments with firmware versions 5 00 or higher use open source software internally Some of this sof...

Страница 11: ...mal single phase supply CONNECTION TO THE POWER SUPPLY An IEC 320 three pronged recessed male socket is provided on the back panel for AC power connection to the instrument One end of the detachable p...

Страница 12: ...erators are prevented access and or are insulated from every connection point In some cases connections must be exposed to potential human contact Operators must be trained to protect themselves from...

Страница 13: ...C decreasing linearly to 50 relative humidity at 40 C 5 Mains supply voltage fluctuations up to 10 of the nominal voltage 6 no pollution or only dry non conductive pollution LABVIEW DRIVERS A LabVIEW...

Страница 14: ...the fuses on the rear panel are shown in the figure below AC FUSE REPLACEMENT To physically access the AC fuses the power cord must be detached from the rear panel of the instrument The fuse drawer ma...

Страница 15: ...Nominal Mains Voltage Rating Case Size Recommended Replacement Part Littelfuse Part Number Digi Key Stock Number 1 2 AC 100 240V 0 5A 250V Time Delay 5 20 mm 0218 500HXP F2416 ND 3 DC N A 1 6A 250V Ti...

Страница 16: ...ed for about 5 seconds after which the instrument will attempt to re enable the output i e light green for about 1 second If the overload condition persists the output will be disabled again i e light...

Страница 17: ...is decreases the setting by a factor of ten If one of the adjustable numeric parameters is displayed and this parameter can be both positive or negative this changes the sign of the parameter EXTRA FI...

Страница 18: ...f the instrument This input can be either active high or active low depending on the front panel settings or programming commands The instrument triggers normally when this input is unconnected When s...

Страница 19: ...It may be necessary to send a few return characters before attempting a login in order to provide enough data to allow this auto sensing to work A standard Linux agetty process is used to implement se...

Страница 20: ...set to a positive value the SYNC pulse precedes the OUT pulse When the delay is set to a negative value the SYNC pulse follows the OUT pulse These pulses are illustrated below assuming internal trigg...

Страница 21: ...ment controls the trigger frequency and generates the clock internally External Trigger the instrument is triggered by an external TTL level clock on the back panel TRIG connector Manual Trigger the i...

Страница 22: ...connector The instrument can be set to stop triggering when this input high or low using the front panel gate menu or the appropriate programming commands When gated the output will complete the full...

Страница 23: ...voltage to the rear panel OS connector Internally it is connected to the output centre conductor via a high quality RF inductor Do not exceed 50V 250 mA ELECTRONIC AMPLITUDE CONTROL EA OPTION The out...

Страница 24: ...may be damaged if triggered by an external signal at a pulse repetition frequency greater than 200 kHz 20 kHz for units with the W2 or W3 options USE A 50 LOAD A 50 Ohm load is required for proper ope...

Страница 25: ...oper operation The output stage may be damaged if the output is not terminated into a 50 load The AVR E1 B output may be observed using either a high speed sampling oscilloscope or a slower real time...

Страница 26: ...some sampling oscilloscopes a 20 dB attenuator should be placed at the input to the scope trigger channel d The bandwidth capability of components and instruments used to display the pulse generator...

Страница 27: ...ow pointer should be pointing at the Normal choice If it is not press MOVE until it is d Press CHANGE to return to the main menu 5 To set the pulse width to 100 ns a Press the MOVE button until the ar...

Страница 28: ...ting to a value more compatible with your sampling oscilloscope Repeat step 4 if required You may also need to adjust the sampling scope controls d Rotate the ADJUST knob The amplitude as seen on the...

Страница 29: ...turns on the output For triggering a single event this sequence would be more appropriate rst resets the instrument trigger source hold turns off all triggering pulse width 50 ns sets the pulse width...

Страница 30: ...value DCYCle numeric value HOLD WIDTh DCYCle DELay numeric value GATE TYPE ASYNC SYNC LEVel HIgh LOw SOURce VOLTage LEVel IMMediate AMPLitude numeric value EXTernal PROTection TRIPped query only STAT...

Страница 31: ...Nual HOLD IMMediate CLS no query form ESE numeric value ESR query only IDN query only OPC SAV 0 1 2 3 no query form RCL 0 1 2 3 no query form RST no query form SRE numeric value STB query only TST que...

Страница 32: ...factory for repair Service is to be performed solely by qualified service personnel Caution High voltages are present inside the instrument during normal operation Do not operate the instrument with t...

Страница 33: ...s recommended TRIGGER DAMAGE The rear panel TRIG input used in the external trigger mode is protected by a diode clamping circuit However the protection circuit is not foolproof and it is possible for...

Страница 34: ...ADBOARD B UNITS ONLY Chassisgroundpost G1 G2 B1 G4 G3 L N G X2 CORCOM6EGG1 2 POWERENTRY M ODULE 1 1b 1a 2 2b 2a X1 POWERSWITCH SW325 ND CW INDUSTRIES GRS 4022 0013 A1 BROWN A2 BLUE A3 BLACK A4 WHITE G...

Страница 35: ...PCB 158R4 LOW VOLTAGE POWER SUPPLY...

Страница 36: ...PCB 235B HIGH VOLTAGE DC POWER SUPPLY...

Страница 37: ...PCB 151D RANGE POLARITY RELAY BOARD...

Страница 38: ...p c b 2 1 7 d d b D o c u m e n ts p c b 2 1 7 a s c h S S R R E L A Y D R IV E R 1 A 1 2 3 4 5 6 K 1 P S 7 1 0 A 1 A 1 2 3 4 5 6 K 2 P S 7 1 0 A 1 A R 2 1 5 K R 1 1 5 K 1 5 V I N 1 2 4 V O U T 2 1 5...

Страница 39: ...MON SCH POLARITY IN TTLTO PG P TTLTO PG N TTLIN PG N 15V PG P 15V V TO PG V TO PG UV ON RELAY POL 15V 15V VCC GND POUT AMP 15V 15V VCC GND RELAY POL UV ON V TO PG V TO PG PCB96F MAIN SCH PCB96F MAIN...

Страница 40: ...33K VCC 9 10 8 U7C MC74F86 VCC DISABLE DISABLE 7 5 Q2B MPQ2222 VCC Q GOES HIGH IF V 15V STAYS HIGH UNTIL VFALLS TO 5V I E HYSTERESIS Q GOES HIGH IF V 15V STAYS HIGH UNTIL V FALLS TO 5V I E HYSTERESIS...

Страница 41: ...5 22K GNDF GNDF 24VF 15VF GNDF 15VF 15VF GNDF GNDF GNDF 15V 15V VCC GND 15V 15V VCC THETRANSISTORS DISCHARGEVc WHENImax EXCEEDED DUE TO CAPACITOR CHARGING OR Vout IS DISABLED DUE TO POLARITYCHANGE UV...

Страница 42: ...2015 ddb Documents Panelbrd prj PANELTOP LEVELSCHEMATIC 1 2 3 4 5 6 7 8 9 10 J5 Amp 54999 10 1 10 pin straight head er I2C_INT SC L SDA GND VC C SINGLE PULSE BACKLIGHT ENCODER ENCODER S CH I2C_INT SC...

Страница 43: ...UST HAVE A IN P N VC C 1A 1B X2 82 101 71 1 b utton keypad 1 2 U4 A MM74HC 14N 11 10 U1 E MM74HC 14N SINGLE PULSE VC C VC C C1 0 2 2uF C1 1 2 2uF C1 2 2 2uF C9 2 2uF C7 2 2uF C6 2 2uF C1 2 2uF 3 4 U4...

Страница 44: ...DA SC L VC C VC C GND VC C VC C C5 0 1uF 1 2 U1 A MM74HC 14N 13 12 U1 F MM74HC 14N 5 6 U1 C MM74HC 14N PAD3 LED PAD4 LED X3 4 40 MOUNT X1 4 40 MOUNT X9 4 40 MOUNT X8 4 40 MOUNT 3 4 U1 B MM74HC 14N LC...

Страница 45: ...6 5VON OFF 27 15V ON OFF 28 O UT SYNC VSPARE VPRF GATE 24 V GND TR IG AUX GND VF NO C NC SW NO C NC AMP LAN O LIM EX 295E BD1 PCB295E CONN2 G ATE CONN3 EXTTRIG CONN1 SYNC 24V NOOLO C 9 3300 pF 20 kHz...

Страница 46: ...PERFORMANCE CHECKSHEET 46...

Отзывы: